[Intra-operative drop in sevoflurane and oxygen concentrations. Leakage of the ventilation system]
M Roiss1, K Bremer, G N Schmidt
1Zentrum für Anästhesiologie und Intensivmedizin, Klinik und Poliklinik für Anästhesiologie, Universitätsklinikum Hamburg-Eppendorf, 20246 Hamburg. m.roiss@uke.uni-hamburg.de
Abstract:
After problem-free induction of narcosis in an 84-year-old female patient an intra-operative drop in sevoflurane and oxygen concentrations occurred during low-flow anesthesia. Although the concentrations of sevoflurane and oxygen in the fresh gas flow were increased no adequate elevation of the inspiratory concentrations could be achieved. Disconnection of the Dräger Primus IE manual bag-valve-mask could be identified as the cause of the drop in concentrations. Interestingly no error alarm function was initiated. This case demonstrates how important knowledge of the function, set-up and alarm conditions of respiratory machines is. This should be an important component of training in anesthesiology as well as securely established algorithms for difficult ventilation to ensure safe anesthesia despite technical failures.
Related Concept Videos
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Inhalational Anesthetics: Overview
Acute Respiratory Failure-II
The underlying physiological abnormalities that contribute to hypoxemic respiratory failure include:
Endotracheal Tube Extubation
Procedure
Extubation removes the endotracheal tube (ETT) from the patient on mechanical ventilation. It requires a well-coordinated, multidisciplinary approach involving physicians, nurses, respiratory therapists, and other healthcare professionals.
Acute Respiratory Failure-I
Definition: It is defined by specific criteria based on blood gas measurements. Hypoxemia happens when the partial pressure of oxygen (PaO2) falls below 60 mmHg. At the same time,...
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...