Direct Pressure Monitoring Accurately Predicts Pulmonary Vein Occlusion During Cryoballoon Ablation
Published on: February 26, 2013
A novel cryoballoon technique for mapping and isolating pulmonary veins: a feasibility and efficacy study
Min Tang1, Charalampos Kriatselis, Sotiris Nedios
1Department of Internal Medicine/Cardiology, German Heart Institute Berlin, Berlin, Germany.
Introduction:
The study was designed to evaluate the feasibility and efficacy of a simplified cryoballoon technique in which a microcircular catheter was introduced into the central lumen of a cryoballoon catheter for the purpose of recording pulmonary vein (PV) potentials during ablation procedures and without interchanging catheters.
Methods And Results:
A total of 23 consecutive patients with paroxysmal atrial fibrillation (AF) were enrolled. A single transseptal puncture was made and a cryoballoon catheter was inserted into the left atrium. A 6-pole mapping catheter with a 0.035-inch shaft diameter was introduced into the PV through the central lumen of the cryoballoon catheter. In addition to the function as a recording device, the mapping catheter was also used as a "guide-wire" during the procedure. A total of 84 PVs (84/92, 91.3%) were completely isolated using this novel cryoballoon technique. In 43 of the 84 veins (51.2%), isolation was observed in real time during the cryoablation; in the remaining 41 veins (48.8%), isolation was confirmed immediately post ablation attempt with the mapping catheter. Procedure time was 152.7 +/- 54.9 minutes, and fluoroscopy time was 33.2 +/- 17.3 minutes. At follow-up (7.4 months, range 2-18 months), 17 (73.9%) patients were free from AF. There was 1 occurrence of phrenic nerve palsy during ablation of a right superior PV, which fully resolved after 1 month.
Conclusion:
The use of a cryoballoon catheter equipped with a 6-pole micromapping catheter inserted through its central lumen for the purpose of mapping and ablation during PV isolation procedures is both feasible and effective.