Saving dose in triple-rule-out computed tomography examination using a high-pitch dual spiral technique
Wieland H Sommer1, Jan C Schenzle, Christoph R Becker
1Departments of Clinical Radiology and daggerRadiotherapy and Radiooncology, University Hospitals-Grosshadern, Ludwig-Maximilians University, Munich, Germany. wieland.sommer@med.uni-muenchen.de
Objective:
High radiation doses remain a drawback of current triple-rule-out computed tomography (CT) protocols. With dual source CT, a new high-pitch dual spiral technique offers the possibility to acquire an Electrocardiography (ECG)-gated-synchronized dataset of the whole chest in less than 1 second. The aim of this study was to compare the dose of such a protocol to a standard, nongated chest scan and to a conventional, retrospectively ECG-gated triple-rule-out protocol. Also, the efficacy and dose of this dual spiral protocol was to be compared in patients examined with this high-pitch technique and matched controls scanned with the conventional technique.
Materials And Methods:
An anthropomorphic Alderson Rando phantom was equipped with thermoluminescent detectors and scanned with the high-pitch protocol (Siemens Somatom Definition Flash; 2 x 120 kVp, 426 mAseff, 128 x 0.6 mm collimation, pitch 3.2), the nongated chest scan (same scanner; 120 kVp, 160 mAseff, 128 x 0.6 mm, pitch 1.2; equivalent Computed Tomography Dose Index (CTDI) of 7.12 mGy), and the conventional gating technique (Siemens Somatom Definition; 120 kVp, 560 mAseff with ECG pulsing interval at 30%-70% of the R-R cycle, 64 x 0.6 mm, pitch 0.3). Noise was measured in air, central and peripheral soft tissue of the phantom. Conversion factors were determined based on the measured dose and the dose-length products of the scanner. The protocol was then applied with ethics committee approval in 31 patients suffering from acute chest pain. The 120 mL of contrast material (Ultravist 370, Bayer Schering Pharma) was applied at 5 mL/s. Dose was calculated based on the dose-length products and the conversion factor. Image quality was assessed by 2 readers for aorta, pulmonary arteries, and coronary arteries. The results were compared with matched controls scanned with the conventional ECG gating technique and non-ECG gated thorax scans.
Results:
The dose determined with thermoluminescent dosimeters measurements amounted to 2.65, 2.68, and 19.27 mSv, respectively, for the dual spiral technique, the standard chest scan, and the conventional retrospective technique. There was no significant difference in image noise. Respective conversion factors were 0.0186, 0.0188, and 0.0180 mSv/mGy x cm. In the patient examinations, dose was 4.08 +/- 0.81 mSv with the high-pitch protocol compared with 20.4 +/- 5.3 mSv in the matched controls with the conventional technique, and 4.40 +/- 0.83 mSv for the non-ECG gated thorax scan. Scan times were 0.7 +/- 0.1 seconds for the high-pitch scan and 15 +/- 3 seconds for the conventional chest pain scan. Aorta and pulmonary arteries were depicted in diagnostic quality in both groups. About 84.7% of coronary artery segments were rated as diagnostic in the high-pitch exams (95.4% below 65 bpm and only 72.8% in higher heart rates), whereas 92.9% were diagnostic with the conventional approach.
Conclusion:
The high-pitch dual spiral technique requires only about one-fifth of the dose of conventional ECG gated triple-rule-out protocols, thus matching that of a standard nongated chest scan. With less than 1 second, the scan time is very short. This protocol can be recommended for patients with unclear chest pain with rhythmic heart rates below 65 bpm.
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies for Cardiovascular System V: CT
Imaging Studies III: Computed Tomography
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...