Related Experiment Video
Updated: Jun 17, 2026

Rejection of Fluorescence Background in Resonance and Spontaneous Raman Microspectroscopy
Published on: May 18, 2011
Vertical double-pass multiple reflection element for internal reflection spectroscopy
1Philips Laboratories, Briarcliff Manor,New York, USA.
Abstract:
A vertical double-pass internal reflection element for internal reflection spectroscopy is described which can be dipped into vessels containing liquids or powders to record their spectra. It has the advantage over the horizontal double-pass internal reflection element in that no additional optical components are required to reorient the spectrometer slit image. Photographs are presented of internal reflection elements showing the actual paths of the light beams outside and inside the internal reflection elements.
Related Concept Videos
IR Spectrometers
Attenuated Total Reflectance (ATR) Infrared Spectroscopy: Overview
The ATR process begins by directing a beam...
UV–Vis Spectrometers
Mass Analyzers: Common Types
Total Internal Reflection Fluorescence Microscopy
Double Resonance Techniques: Overview
Spin decoupling is usually achieved by...