The calibration of scanning monochromators for the measurement of absolute irradiance

J J Merrill1, R Gary Layton

  • 1Physics Department, Utah State University, Logan, Utah, USA.

Applied Optics
|January 9, 2010
PubMed

Related Concept Videos

Instrument Calibration01:12

Instrument Calibration

Instrument calibration is essential for ensuring that instruments produce accurate and consistent results. It is vital in manufacturing, healthcare, testing laboratories, and scientific research. Calibration processes are specific to each instrument and help enhance data accuracy. Each instrument has a unique calibration process tailored to its design and function to improve data accuracy.
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...
Atomic Absorption Spectroscopy: Instrumentation01:22

Atomic Absorption Spectroscopy: Instrumentation

An atomic absorption spectrophotometer (AAS) comprises several components: a radiation source, an atomizer, a monochromator, and a detector. The radiation source can be a hollow-cathode lamp (HCL) or an electrodeless-discharge lamp (EDL), both of which provide a narrow emission line of the required wavelength. However, some instruments use continuum sources and high-resolution monochromators to achieve a narrow range of radiation.
The atomizer used in AAS can be either a flame atomizer or an...
UV–Vis Spectrometers01:14

UV–Vis Spectrometers

The absorbance of UV and visible (UV–visible) radiations is measured using a UV–visible spectrophotometer. Deuterium lamps, which emit UV radiation, and tungsten lamps, which produce radiation in the visible region, are used as light sources in UV–visible spectrophotometers. A monochromator or prism is used for diffraction grating, i.e., to split the incoming radiation into different wavelengths. A system of slits is used to focus the desired wavelength on the sample cell. Samples for...
IR Spectrometers01:25

IR Spectrometers

There are two main infrared (IR) spectrophotometers: dispersive IR spectrometers and Fourier transform infrared (FTIR) spectrometers. In a dispersive IR spectrometer, a beam of infrared radiation produced by a hot wire is divided into two parallel equal-intensity beams using mirrors. One beam passes through the sample, while another is a reference beam. The beams then move through the monochromator, which separates the radiations into a continuous spectrum of different frequencies. The...
Atomic Absorption Spectroscopy: Radiation and Light Sources01:13

Atomic Absorption Spectroscopy: Radiation and Light Sources

Atomic absorption spectroscopy (AAS) relies on the Beer-Lambert law, which requires that the radiation source emits a narrow range of wavelengths to match the absorption characteristics of the analyte atom. The primary criteria for choosing an appropriate radiation source in AAS is to provide a precise and intense emission at specific wavelengths that will allow accurate detection of the analyte.
Two common narrow-range 'line' sources used in AAS are hollow-cathode lamps (HCLs) and...