Characteristics of telemetry interference with pacemakers caused by digital media players
Ashok J Shah1, Joseph D Brunett, Jay P Thaker
1Thoracic and Cardiovascular Institute, Sparrow Health System, Michigan State University, Lansing, Michigan 48910, USA.
Background:
Contemporary implantable heart rhythm devices communicate multiple complex data simultaneously using radiofrequency telemetry. Interference in communication can expose them to the risk of potential corruption, leading to adverse clinical consequences.
Methods & Results:
We studied the characteristics of interference with uplink (real time intracardiac electrograms, marker channel, and stored histograms) and downlink (attempt to program a change in the lower rate limit, the pacing mode, and the ventricular lead configuration) data transmission between the wand and the pacemaker caused by digital media players (iPods--Photo and 3G) in 50 patients. We also measured and characterized worst-case magnetic field emissions (MFE) from the wand (
Conclusion:
Low-power MFE from iPods can produce interference with establishment and maintenance of a telemetry link and can cause TI with transmission of real time data, but because of continuous check protocols, do not corrupt the stored and vital downlink data.
Related Concept Videos
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac muscle...
Holter Monitor: 24-Hour Monitoring
Instrumentation Amplifier
To overcome this challenge, an ECG machine utilizes an instrumentation amplifier. This specialized amplifier is...
Disturbances in Heart Rhythm
Arrhythmias are categorized by their speed, rhythm, and origin. A slow heart...