Using Reference Reagents to Confirm Robustness of Cytokine Release Assays for the Prediction of Monoclonal Antibody Safety
Published on: September 15, 2023
Performance characteristics of three random access cyclic citrullinated peptide antibody assays
Sara P Wyness1, Sonia L La'ulu, Joshua J H Hunsaker
1ARUP Institute for Clinical and Experimental Pathology, Salt Lake City, UT, USA.
Background:
Cyclic citrullinated peptide antibodies (CCP Ab) are useful biomarkers for the early detection and diagnosis of rheumatoid arthritis (RA).
Methods:
We evaluated the performance of 3 random access 2nd generation CCP Ab assays, the Abbott AxSYM and Architect analyzers and the Roche Modular Analytics E170 analyzer, for limit of detection (LOD), imprecision, results for samples from healthy subjects, analytic concordance, and interferences. Method comparison testing was performed using the AxSYM analyzer as the comparison method and a 3rd generation INOVA Quanta Lite ELISA assay was included.
Results:
LOD determinations met the manufacturers' claims. Total CVs ranged from 1.6% to 8.2%. Results from healthy subjects were generally much lower than the manufacturers' decision cutoffs. Comparison to the AxSYM assay resulted in overall concordance ranging from 82.1% to 98.3%. These assays were resistant to interference from hemolysis, icterus, lipemia and rheumatoid factor.
Conclusion:
All 3 random access CCP Ab assays performed according to the manufacturers' claims and have the potential to improve workflow in clinical laboratories.