Related Experiment Video
Updated: Jun 17, 2026

Large Scale Non-targeted Metabolomic Profiling of Serum by Ultra Performance Liquid Chromatography-Mass Spectrometry (UPLC-MS)
Published on: March 14, 2013
Development and validation of high-throughput liquid chromatography-tandem mass spectrometric method for simultaneous
Raghunadha Reddy S1, Koteswara Rao Divi, I Sarath Chandiran
1Clinical Research and Biosciences (I) Pvt. Ltd., Hyderabad, Andhra Pradesh, India. raghu_nathseelam@yahoo.co.in
Abstract:
A simple, sensitive and reliable method is described for simultaneous quantification of Clopidogrel and its metabolite in human plasma by using HTLC-MS/MS. The analytical procedure involves on-line coupling of extraction with Cyclone P (50 mm x 0.5 mm 50 microm) HTLC column by injecting 15 microL sample and chromatographic separation is performed with Cohesive Propel C18 (5 microm, 3.0 mm x 50 mm), followed by quantification with mass detector in SRM mode using ESI as an interface. The calibration curves were linear over a concentration range of 0.1-8 ng/mL of Clopidogrel and 70 ng/mL to 6 microg/mL of its metabolite using 20 mL human plasma per batch. The total run time of analysis was 7.5 min and the lower limits of quantification were 0.1 ng/mL for Clopidogrel and 70 ng/mL for its metabolite. The method validation was carried out in terms of specificity, sensitivity, linearity, precision, accuracy and stability. The validated method was applied in bioavailability and bioequivalence study.