Related Experiment Video
Updated: Jun 17, 2026

Angle-resolved Photoemission Spectroscopy At Ultra-low Temperatures
Published on: October 9, 2012
La pulvérisation par ultra-sons appliquée à la spectrometrie d'absorption atomique
Abstract:
Ultrasonic nebulization applied to atomic absorption spectrometry gives a sensitivity ten to twenty times better than that which can be obtained by classical pneumatic nebulization. This gain in sensitivity is primarily due to the quantity of aerosol introduced into the flame, but is equally due to the fineness of the droplets that are formed. The study of certain parameters (frequency, power applied to the emitter, geometry of the nebulizing chamber, gas flow) has led to the specification of an ultrasonic nebulizer that can be conveniently used for routine analyses.
Related Concept Videos
Atomic Absorption Spectroscopy: Atomization Methods
Atomic Spectroscopy: Absorption, Emission, and Fluorescence
Atomic Absorption Spectroscopy: Instrumentation
The atomizer used in AAS can be either a flame atomizer or an...
Atomic Absorption Spectroscopy: Lab
Solutions containing organic solvents, such as low-molecular-mass alcohols, esters, or ketones, enhance absorbances by increasing nebulizer...
Atomic Emission Spectroscopy: Instrumentation
Atomic Absorption Spectroscopy: Overview
When irradiated by EMR of a particular wavelength, these...