Related Experiment Video
Updated: Jun 16, 2026

Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
3D Dosimetric verification of volumetric-modulated arc therapy by portal dosimetry
Anton Mans1, Peter Remeijer, Igor Olaciregui-Ruiz
1Department of Radiation Oncology, The Netherlands Cancer Institute-Antoni van Leeuwenhoek Hospital, Amsterdam, The Netherlands. a.mans@nki.nl
Background And Purpose:
To demonstrate the feasibility of back-projection portal dosimetry for accurate 3D dosimetric verification of volumetric-modulated arc therapy (VMAT), pre-treatment as well as in vivo.
Materials And Methods:
Several modifications to our existing approach were implemented to make the method applicable to VMAT: (i) gantry angle-resolved data acquisition, (ii) calculation of the patient transmission, (iii) compensation for detector 'flex' and (iv) 3D dose reconstruction and evaluation.
Results:
Planned and EPID-(Electronic Portal Image Detector)-reconstructed dose distributions show good agreement for pre-treatment verification of two prostate, a stereotactic lung and a head-and-neck VMAT plan and for in vivo verification of VMAT treatments of prostate and lung cancer. Averaged over pre-treatment verifications, planned and measured isocentre dose ratios were -1.2% (range [-4.7%,1.8%]). 3D gamma analysis (3% maximum dose, 3mm) revealed mean gamma gamma(mean)=0.37 [0.34,0.39], maximum 1% gamma gamma(1%)=0.72 [0.66,0.81] and percentage of points with gamma1 P(gamma)(1)=99% [97%,100%]. For in vivo verification, the average isocentre dose ratio was -1.2% [-0.8%,-1.7%], gamma(mean)=0.52 [0.40,0.64], gamma(1%)=0.92 [0.76,1.08] and P(gamma)(1)=96% [93%,100%].
Conclusions:
Our portal dosimetry method was successfully adapted for verification of VMAT treatments, pre-treatment as well as in vivo.