Cardiac output derived from arterial pressure waveform analysis without calibration vs. thermodilution in septic
Cornelis Slagt1, Jan Beute, Martijn Hoeksema
1Department of Anaesthesiology, Intensive Care and Pain Relief, Zaans Medical Centre, Zaandam, The Netherlands. Slagt.c@zaansmc.nl
Background And Objective:
We studied the evolution of software in the accuracy of the FloTrac/Vigileo system to measure cardiac output less invasively from arterial pressure waveform analysis without calibration, in comparison with pulmonary artery catheter-derived thermodilution measurements, in patients with septic shock and presumed alterations in vascular tone.
Methods:
Nine patients who received a pulmonary artery catheter and were on mechanical ventilation and in sinus rhythm were monitored by the FloTrac/Vigileo. Paired cardiac output measurements by both techniques were analysed for 86 measurements in four patients using the 1.07 software version and 73 measurements in five subsequent patients using the later 1.10 version.
Results:
For the 1.07 version, bias was -1.6 L min, precision 1.6 L min, limits of agreement -4.8-1.5 L min and error 48%. Measurements correlated at partial r equal to 0.32 (P = 0.003). For the 1.10 version, bias was -1.2 L min, precision 1.1 L min, limits of agreement -3.5-1.0 L min and error 32%. Measurements correlated at partial r equal to 0.90 (P < 0.001 vs. version 1.07). Differences were inversely related to mean cardiac output (P < 0.001, generalized estimating equations), particularly for software version 1.07 vs. 1.10 (P = 0.017, generalized estimating equation). Changes in thermodilution cardiac output over the course of time were also better tracked by the FloTrac/Vigileo when applying the latest software (P < 0.001, generalized estimating equation).
Conclusions:
Evolving software versions are thus better able to account for the effect of vascular tone on cardiac output measurements by less invasive waveform analyses without calibration (FloTrac/Vigileo), so that the latter may become useful in the haemodynamic monitoring of septic shock.
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Measurement of Blood Pressure
Cardiac Output and Stroke Volume
In an average resting adult male, the typical cardiac output averages...
Cardiac Output II: Effect of Stroke Volume on Cardiac Output
Preload
Preload refers to the initial elongation of the cardiac myocytes before contraction and is related to the volume of blood filling the heart at the end of diastole, or end-diastolic volume. The...