Biomechanical testing of two devices for internal fixation of fractured ribs
Nicole Campbell1, Martin Richardson, Phillip Antippa
1Department of Cardiothoracic Surgery, Royal Melbourne Hospital, Parkville, Victoria 3052, Australia. nicole.campbell@mh.org.au
Background:
An intramedullary screw (Portland Orthopaedics, St. Clair, MI) and Inion (Tampere, Finland) Orthopaedic Trauma Plating System (OTPS) mesh for use for internal fixation of fractured ribs have not previously undergone biomechanical testing. The aim of this study is to compare the biomechanics of intact ribs undergoing four-point bending to the biomechanics of fractured ribs fixed with each of the two devices to determine which device provides superior fixation.
Study:
Thirty fresh-frozen porcine ribs (ribs 6-8) were submitted to four-point bending to failure, at a rate of 2.5 mm/min, to determine stiffness and force at failure. The ribs were then randomized to receive fixation with either Inion OTPS mesh or an intramedullary screw. The fixated ribs were again submitted to four-point bending to failure.
Results:
Ribs fixated with OTPS mesh were significantly stiffer and failed at a significantly higher load than ribs fixated with intramedullary screws (p = 0.0001). Ribs fixated with OTPS mesh were less stiff than intact ribs but failed at a similar force to intact ribs.
Conclusion:
The Inion OTPS mesh system provides superior fixation of fractured ribs compared with an intramedullary screw when tested with four-point bending. The OTPS mesh system also has the added advantage of being absorbable. The intramedullary screw needs to undergo modification and further testing before use in humans. Future studies should focus on more physiologic methods of loading.
More Related Videos
11:21Development of a Novel Internal Fixation Model for Rat Radial Fractures: Fracture Healing Assessment and Dorsal Root Ganglion Isolation
Published on: March 13, 2026
06:41A Minimally Invasive Model to Analyze Endochondral Fracture Healing in Mice Under Standardized Biomechanical Conditions
Published on: March 22, 2018
Related Concept Videos
Flail Chest-II
Assessment:
1. Clinical Evaluation:
History:
Fractures: Bone Repair
Minor fractures with no bone displacement are treated by immobilizing the fractured bone using a cast or splint. However, in the case of fractures with displaced bones, the broken bones are repositioned before immobilization to ensure successful healing without deformation and loss of function. The realignment of fractured bone ends is performed through a process called reduction. If the procedure...