Related Experiment Video
Updated: Jun 16, 2026

Proton Therapy Delivery and Its Clinical Application in Select Solid Tumor Malignancies
Published on: February 6, 2019
Measurement of neutron dose equivalent and its dependence on beam configuration for a passive scattering proton
Xin Wang1, Narayan Sahoo, Ronald X Zhu
1Department of Radiation Physics, The University of Texas M. D. Anderson Cancer Center, Houston, TX 77030, USA.
Purpose:
To measure the neutron dose equivalent per therapeutic proton dose (H/D) in a passive scattering proton therapy system and study its dependence on the proton energy, aperture-to-isocenter distance, spread-out Bragg peak (SOBP) width, and field size.
Methods And Materials:
We performed four experiments of varying proton energies, aperture-to-isocenter distances, SOBP widths, and field sizes. Etched track detectors were used to measure the neutron dose equivalent at both an in-field (isocenter, beyond the protons' range) and out-of-field (30 cm lateral to the isocenter) location in air.
Results:
For a nonmodulated beam with all the protons stopping in the aperture and an aperture-to-isocenter distance of 30 cm, the H/D values measured at the isocenter were approximately 0.3 mSv/Gy for all snouts with a 100-MeV beam. The H/D values increased to 10.7, 14.5, and 15.1 mSv/Gy, respectively, for small, medium, and large snouts when the beam energy increased to 250 MeV. At the out-of-field location, H/D values increased from 0.1 to 2.7, 3.0, and 3.2 mSv/Gy, respectively, for small, medium, and large snouts. When the aperture-to-isocenter distance was changed from 10 to 40 cm, the H/D value at the isocenter dropped 70%. The H/D value doubled for the modulated beam relative to the nonmodulated beam. Open apertures reduced the neutrons produced in the nozzle, but increased those produced in the phantom.
Conclusions:
Our data showed that changes in the four factors studied affect the H/D value in predictable ways which permits an estimate of a patient's neutron exposure.
More Related Videos
10:24Neutron Radiography and Computed Tomography of Biological Systems at the Oak Ridge National Laboratory's High Flux Isotope Reactor
Published on: May 7, 2021
07:31Characterization of Recombination Effects in a Liquid Ionization Chamber Used for the Dosimetry of a Radiosurgical Accelerator
Published on: May 9, 2014
Related Concept Videos
Biological Effects of Radiation
Types of Radioactivity
Alpha (α) decay is the emission of an α particle from the nucleus. For example, polonium-210 undergoes α decay:
Nuclear Stability
To hold positively charged protons together in the...