Concept of minimal heart rate for each pitch value to avoid interpolation artifact when using dual-source CT: a
Joon-Won Kang1, Kyung-Hyun Do, Jae-Yeon Chung
1Department of Radiology and Research Institute of Radiology, University of Ulsan College of Medicine, Asan Medical Center, 388-1 Pungnap-2-dong, Songpa-gu, Seoul, 138-736, Republic of Korea.
Abstract:
Interpolation artifact is known to occur when the heart rate is decreased lower than the critical value for the specific pitch. The purpose of our study is to determine the minimum heart rate (minHR) for the specific pitch that provides images without interpolation artifact when using dual-source computed tomography (DSCT). We scanned the 'thin slice thickness block' of the CT performance phantom provided by the American Association of Physicists in Medicine using DSCT for variable pitches. Change in heart rate was simulated through ECG editing by changing R-R interval. Axial, sagittal, and coronal image sets were reconstructed and assessed for the presence and extent of interpolation artifact. MinHR at which no interpolation artifact was detected for each pitch value was determined. Length of interpolation artifact (LOA) on sagittal view was also measured when the heart rate was simulated at 10 bpm lower than the minHR on each pitch setting. MinHRs for each pitch value were 9-10 bpm from the estimated heart rate. However, minHR for the lowest pitch value 0.2, estimated heart value 40 bpm was 37 bpm. LOA was larger in the low heart rate condition. Measured values of minHR were correlated exactly with the calculated values. MinHRs that provide images without interpolation artifact for each pitch value when using DSCT were determined. The concept of minHR is important for obtaining high quality images of cardiac CT angiography when using DSCT.
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Cardiac Output I:Effect of Heart Rate on Cardiac Output
Cardiac output (CO) refers to the total amount of blood ejected by one of the ventricles in liters per minute (L/min). In a resting adult, CO ranges from 5 to 6 L/min, adjusting according to the body's metabolic requirements.
Effect of Heart Rate on Cardiac Output
Cardiac output adapts to metabolic demands during stress, physical activity, or illness. The autonomic nervous system regulates heart rate via the sinoatrial node. The parasympathetic nervous system decreases heart rate...
Special considerations while measuring pulse