Catheter-based Endovascular Angioplasty for Fibrosing Mediastinitis-associated Pulmonary Vein Stenosis
Published on: August 26, 2025
Multiple coaxial catheter system for reliable access in interventional stroke therapy.
Zsolt Kulcsár1, Hasan Yilmaz, Christophe Bonvin
1Neurointerventional Division, Department of Clinical Neurosciences, Geneva University Hospital, Geneva, Switzerland. kulcsarzsolt22@gmail.com
Navigating tortuous vessels during mechanical thrombectomy for acute ischemic stroke is challenging. A novel coaxial catheter system demonstrated safety and efficiency in improving access and stability for revascularization procedures.
Area of Science:
- Neurology
- Interventional Neuroradiology
- Medical Devices
Background:
- Navigating complex aortic arch and cerebral vasculature is a significant challenge during mechanical thrombectomy for acute ischemic stroke.
- Existing thrombectomy systems can face limitations in tortuous anatomy, potentially hindering successful revascularization.
Purpose of the Study:
- To describe a novel multiple coaxial catheter system designed to facilitate navigation and manipulation in tortuous vessels.
- To evaluate the safety and efficacy of this system in patients undergoing mechanical thrombectomy for acute ischemic stroke.
Main Methods:
- A triple or quadruple coaxial catheter system was utilized in 28 consecutive acute ischemic stroke cases.
- The system comprised 8-Fr and 6-Fr guiding catheters housing inner reperfusion catheters, navigated over a microcatheter.
- Mechanical thrombectomy was performed using the Penumbra System, with adjunct thrombolysis or angioplasty/stenting as needed.
Main Results:
- The coaxial catheter system successfully provided access to cerebral lesions and stability for complex manipulations, including thromboaspiration and stent navigation in all cases.
- The system enabled diverse interventions, such as carotid stenting and intracranial procedures.
- No complications related to the catheter system were reported in this series.
Conclusions:
- The developed triple or quadruple coaxial catheter system is safe and efficient for mechanical thrombectomy in acute ischemic stroke.
- This system offers improved navigability and stability in tortuous anatomy, enhancing treatment delivery.
- The system's versatility supports various interventional approaches beyond simple thrombectomy.