Anesthetic agent vapor analyzers and propellants of pressurized meter-dose inhalers
Kavaraganahalli Mukundarao Deepak1, Kiran Kini
1Kasturba Medical College, Manipal, India. drdeepakkm@yahoo.co.in
Abstract:
Anesthetic agent analyzers fail when drug from a HFA propelled pMDI is administered. Most anesthetic vapor analyzers use infrared absorption at 3.3 micron. At this wavelength, substance like methane has been reported to interfere with the accuracy of measurement of anesthetic vapor. The anesthetic gas monitor 1304 (Brüel and Kjaer) which functions at 10.3-13 micron wavelength was not affected by methane. Is it possible, HFA with its structural similarity to inhaled anesthetic agents may be responsible for faulty reading of anesthetic vapor concentration in two of our monitors? Further evidence is needed to support this finding. Anesthesiologists need to be ever vigilant and recognize the need for smarter designs of anesthetic agent analyzers with changing array of drugs administered during anesthetic management.
More Related Videos
06:57Use of an Integrated Low-Flow Anesthetic Vaporizer, Ventilator, and Physiological Monitoring System for Rodents
Published on: July 9, 2020
09:36Halogenated Agent Delivery in Porcine Model of Acute Respiratory Distress Syndrome via an Intensive Care Unit Type Device
Published on: September 24, 2020
Related Concept Videos
Inhalational Anesthetics: Overview
Inhaled Medications
Asthma-IV: Diagnostic and Management
Clinical Assessment for Asthma:
This is the first step in diagnosing and managing asthma. It includes:
Parenteral Anesthetics: Overview
Atomic Absorption Spectroscopy: Atomization Methods
Gas Chromatography: Types of Detectors-II