Polyacrylamide gel photopatterning enables automated protein immunoblotting in a two-dimensional microdevice
1Department of Bioengineering, University of California, Berkeley, California 94720, USA.
Abstract:
We demonstrate a two-dimensional microfluidic architecture that integrates polyacrylamide gel electrophoresis (PAGE) with immunoblotting in a fully automated format. This assay is designed to overcome performance limitations of conventional slab-gel immunoblotting, including multiple manual interventions, low-throughput transfer and blotting, and substantial consumption of reagents and sample. To unify PAGE with blotting in one device, this microfluidic approach makes use of high-resolution regional photopatterning of multiple polyacrylamide gel elements, and automated electrophoretic transport. A complete native immunoblot of free prostate specific antigen from human seminal fluid is demonstrated in less than 5 min. Further, the characterization of post-PAGE electrophoretic transfer showed high efficiency and minimal sample dispersion.
Related Concept Videos
Two-dimensional Gel Electrophoresis
The first dimension separation uses the isoelectric focusing or IEF technique performed on immobilized pH gradient (IPG) strips that separate proteins according to their isoelectric points.
Biological samples, such as cells...
SDS-PAGE
A variation of gel electrophoresis, termed polyacrylamide gel electrophoresis (PAGE), is commonly used for separating proteins according to their molecular size by passing them through a polyacrylamide gel. Because of the varying charges associated with amino acid side chains, PAGE can be used to separate intact proteins...