Beware of carryover in modern chemistry analyzers
1Department of Pathology, University of Mississippi Medical Center, Jackson, MS 39232, USA. pkyle@pathology.umsmed.edu
Background:
Random-access analyzers that employ dedicated probes may be affected by carryover. Sample carryover, involving analyte from one sample that is measured in a subsequent sample, is most often detected in measurement procedures that have wide reportable ranges. However, reagent carryover can be more difficult to detect as it may involve only one pair of measurement procedures.
Methods:
Our laboratory noted several patient samples with total cholesterol <2.58 mmol/L (100 mg/dL) during the initial months after a new chemistry analyzer was installed. The problem seemed to occur intermittently, approximately once per week.
Results:
Immediate reanalysis of affected samples resulted in measured values that were 1.03-2.58 mmol/L (40-100 mg/dL) higher. Evaluation of reagent carryover revealed a significant decrease in total cholesterol after analysis of creatine kinase (EC 2.7.3.2). Carryover disappeared when an additional reagent probe wash was applied, whereas the root of the problem was eliminated with replacement of the reagent probes. The insidious nature of reagent carryover made it difficult to initially detect the problem, which affected two of the three instruments in our laboratory.
Conclusions:
Laboratorians should be aware of the potential for carryover in random access analyzers, and should be mindful of appropriate troubleshooting techniques.
More Related Videos
08:56Detection of Regulated Ergot Alkaloids in Food Matrices by Liquid Chromatography-Trapped Ion Mobility Spectrometry-Time-of-Flight Mass Spectrometry
Published on: November 22, 2024
05:48Analysis of Volatile and Oxidation Sensitive Compounds Using a Cold Inlet System and Electron Impact Mass Spectrometry
Published on: September 5, 2014
Related Concept Videos
Inductively Coupled Plasma-Mass Spectrometry (ICP-MS): Interferences
Atomic Absorption Spectroscopy: Interference
Spectral interference occurs when signals from other elements or molecules overlap with the analyte signal, falsely elevating or masking the analyte's absorbance. This interference can be corrected using Zeeman,...
Atomic Absorption Spectroscopy: Lab
Solutions containing organic solvents, such as low-molecular-mass alcohols, esters, or ketones, enhance absorbances by increasing nebulizer...
High-Performance Liquid Chromatography: Types of Detectors
Mass Analyzers: Common Types
Atomic Emission Spectroscopy: Interference