Related Experiment Video
Updated: Jun 16, 2026

10:42
A Postoperative Evaluation Guideline for Computer-Assisted Reconstruction of the Mandible
Published on: January 28, 2020
Computer-assisted surgery in the edentulous jaw based on 3 fixed intraoral reference points
Gerlig Widmann1, Martin Keiler, Antoniette Zangerl
1Stereotaxy Intervention Planning, Department for Microinvasive Therapy, Department of Radiology, Innsbruck Medical University, Innsbruck, Austria. gerlig.widmann@i-med.ac.at
Summary
This study shows that using three fixed intraoral reference points for dental implant surgery in edentulous jaws provides accurate surgical navigation. This method achieves registration and targeting accuracy comparable to traditional techniques.
Area of Science:
- Oral and Maxillofacial Surgery
- Biomedical Engineering
- Dental Implantology
Background:
- Computer-assisted implantation surgery relies on accurate transfer of prosthodontic plans via registration templates.
- Edentulous jaws present challenges for precise registration template repositioning, impacting surgical accuracy.
Purpose of the Study:
- To evaluate the in vitro registration and targeting accuracy of surgical navigation in edentulous jaws.
- To assess the efficacy of using three fixed intraoral reference points for this purpose.
Main Methods:
- Utilized edentulous cadaver maxilla and mandible specimens with three fixed reference-point screws.
- Employed a resin template, registration mouthpiece, and optical navigation system for implant planning and drilling.
- Assessed accuracy through fiducial registration error and post-surgical scan fusion with pre-surgical planning.
Main Results:
- Performed 14 registrations and 104 stone cast drillings across 9 maxillas and 5 mandibles.
- Achieved a mean fiducial registration error of 0.49 ± 0.14 mm.
- Reported mean total error at the borehole tip of 0.88 ± 0.65 mm, with mean lateral errors of 0.51 mm at the base and 0.46 mm at the tip, and a mean angular error of 0.83 ± 0.60 degrees.
Conclusions:
- Three fixed intraoral reference points effectively support a registration mouthpiece for computer-assisted surgery.
- This method demonstrates in vitro registration and targeting accuracy comparable to tooth-supported templates or bone markers.
- Offers a viable solution for improving surgical navigation accuracy in edentulous jaws.