Related Experiment Video
Updated: Jun 16, 2026

Dynamic Electrochemical Measurement of Chloride Ions
Published on: February 5, 2016
Interpretation of chronopotentiometric transients of ion-selective membranes with two transition times
Justin M Zook1, Sándor Bodor, Róbert E Gyurcsányi
1Department of Biomedical Engineering, The University of Memphis, Memphis, TN 38152, USA.
Abstract:
Passing currents through ion-selective membranes has contributed to the development of a variety of novel methods. In this work, chronopotentiometric (CP) transients with two transition times (breakpoints) are presented for the first time, with the theoretical interpretation of such voltage transients. The validity of our theory has been confirmed in experiments utilizing ETH 5294 chromoionophore-based pH sensitive membranes with and without lipophilic background electrolyte and ETH 5234 ionophore-based calcium selective membranes in which the ionophore forms 3:1 complexes with Ca(2+) ions. The conditions under which two breakpoints can be identified in the chronopotentiometric voltage transients are discussed.Spectroelectrochemical microscopy (SpECM) is used to show that the two breakpoints in the CP curves emerge approximately when the free ionophore and ion-ionophore complex concentrations approach zero at the opposite membrane-solution interfaces. The two breakpoint times can be utilized to follow simultaneously the concentration changes of the free ionophore, the ion-ionophore complex, and the mobile anionic sites in cation-selective membranes. In membranes with known composition, the time instances where breakpoints occur can be used to estimate the free ionophore and the ion-ionophore complex diffusion coefficients.
Related Concept Videos
Potentiometry: Membrane Electrodes
Resting Membrane Potential
The Inside of a Neuron is More Negative
The membrane potential of a cell can be measured by inserting a microelectrode into a cell and comparing the charge to a reference electrode in the extracellular fluid. The...
Resting Potential Decay
At rest, the K+ is the main ion that moves across the membrane through...
Dialysis
Transport Number
Potentiometry: Overview