Rejection of Fluorescence Background in Resonance and Spontaneous Raman Microspectroscopy
Published on: May 18, 2011
Premonochromator for stray light rejection in Raman spectroscopy.
Applied Optics
|February 20, 2010
Summary
A novel apparatus effectively attenuates Rayleigh light using dual concave gratings. This innovation enables precise spectral cutting, significantly improving Raman spectroscopy sensitivity for scientific research.
Area of Science:
- Spectroscopy
- Optical Physics
- Instrumentation
Background:
- Rayleigh light scattering interferes with sensitive spectroscopic measurements.
- Accurate detection of Raman signals near the excitation line is challenging.
Purpose of the Study:
- To develop and describe a new apparatus for attenuating Rayleigh light.
- To enable sensitive detection of Raman lines close to the exciting line.
Main Methods:
- Utilized two concave gratings in a nondispersive configuration.
- Employed a movable mirror for sharp spectral cutting at desired positions.
- Integrated the apparatus as a pre-eliminating unit with a double monochromator.
Main Results:
- Achieved Rayleigh line rejection better than 10(-5), optimally 10(-6).
- Enabled measurement of Raman lines as close as 10 cm(-1) from the exciting line.
- Demonstrated successful application as a pre-eliminating unit.
Conclusions:
- The described apparatus offers superior Rayleigh light attenuation.
- This instrument significantly enhances the capability for Raman spectroscopy studies.
- Potential for use with a single monochromator for advanced Raman analysis.
Related Concept Videos
Raman Spectroscopy Instrumentation: Overview
A conventional Raman spectrophotometer includes a laser source, a sample holding system, a wavelength selector, and a detector.
The monochromatic laser source, typically using visible or near-infrared radiation, generates a highly focused beam of light. This light interacts with the molecules of the sample, scattering some of the light. Liquid and gaseous samples are usually tested in ordinary glass capillaries, while solids can be analyzed as powders packed in capillaries or as potassium...
The monochromatic laser source, typically using visible or near-infrared radiation, generates a highly focused beam of light. This light interacts with the molecules of the sample, scattering some of the light. Liquid and gaseous samples are usually tested in ordinary glass capillaries, while solids can be analyzed as powders packed in capillaries or as potassium...
Raman Spectroscopy: Overview
The underlying principle of Raman spectroscopy is based on the interaction between light and matter, specifically molecules' inelastic scattering of photons. When a monochromatic beam of light, typically from a laser source, interacts with a sample, most scattered light has the same frequency as the incident light. This is known as Rayleigh scattering.
However, a small fraction of the scattered light exhibits a frequency shift due to the exchange of energy between the incident photons and the...
However, a small fraction of the scattered light exhibits a frequency shift due to the exchange of energy between the incident photons and the...
IR Spectrometers
There are two main infrared (IR) spectrophotometers: dispersive IR spectrometers and Fourier transform infrared (FTIR) spectrometers. In a dispersive IR spectrometer, a beam of infrared radiation produced by a hot wire is divided into two parallel equal-intensity beams using mirrors. One beam passes through the sample, while another is a reference beam. The beams then move through the monochromator, which separates the radiations into a continuous spectrum of different frequencies. The...
UV–Vis Spectrometers
The absorbance of UV and visible (UV–visible) radiations is measured using a UV–visible spectrophotometer. Deuterium lamps, which emit UV radiation, and tungsten lamps, which produce radiation in the visible region, are used as light sources in UV–visible spectrophotometers. A monochromator or prism is used for diffraction grating, i.e., to split the incoming radiation into different wavelengths. A system of slits is used to focus the desired wavelength on the sample cell. Samples for...