Evaluation of five routine glucose methods on an Olympus AU5400 analyzer using the CDC hexokinase reference method.
1Department of Clinical Laboratory, Peking University Third Hospital, Beijing, China.
Clinical Chemistry and Laboratory Medicine
|February 23, 2010
Summary
Five commercial glucose reagent kits, using glucose oxidase (GOD) and hexokinase (HK) methods, were evaluated against a CDC HK reference method. All kits showed excellent correlation and minimal bias, proving acceptable for clinical laboratory use.
Area of Science:
- Clinical Chemistry
- Biotechnology
- Medical Diagnostics
Background:
- Serum glucose measurement is crucial in clinical diagnostics.
- Two primary methods, glucose oxidase (GOD) and hexokinase (HK), are widely used in China.
- Evaluation of commercial reagent kits against a reference standard is essential for quality control.
Purpose of the Study:
- To assess the performance of five commercially available glucose reagent kits.
- To compare these kits against the Center for Disease Control (CDC) hexokinase (HK) reference method.
- To determine the suitability of these glucose reagents for routine clinical laboratory use.
Main Methods:
- Five routine glucose detection methods were evaluated.
- Comparison was performed using the Clinical and Laboratory Standards Institute (CLSI) guideline EP9-A2.
- Methods were analyzed on an Olympus 5400 analyzer in an open-channel mode.
Main Results:
- All five evaluated glucose reagent kits demonstrated a high correlation (r > 0.9998) with the HK reference method.
- Minor positive biases were observed for Olympus-HK and Roche-GOD methods.
- Slight negative biases were noted for Biosino-HK, Osborne-GOD, and Baiding-GOD methods, all within acceptable clinical limits.
Conclusions:
- The evaluated glucose reagent kits, when used with matching calibrators, exhibit strong correlation and minimal bias compared to the reference method.
- All five commercial glucose reagents are deemed suitable for application in clinical laboratory settings.
- This study validates the reliability of these glucose testing kits for accurate patient diagnosis.