Related Experiment Video
Updated: Jun 16, 2026

06:46
Pulsed Laser Diode-Based Desktop Photoacoustic Tomography for Monitoring Wash-In and Wash-Out of Dye in Rat Cortical Vasculature
Published on: May 30, 2019
Optoacoustic detection of NO(2) using a pulsed dye laser
Applied Optics
|February 23, 2010
Abstract:
Using a pulsed dye laser and a nonresonant optoacoustic detector, the absorption spectrum of NO(2) has been measured in the spectral region of 480-625 nm at various cell pressures and at various concentrations in air. The results of the measurements as a function of NO(2) concentration in air demonstrate the detection capability of the system. An extrapolated sensitivity of 4 ppb NO(2)/W of laser power has been achieved.