Related Experiment Video
Updated: Jun 15, 2026

Measuring 3D In-vivo Shoulder Kinematics using Biplanar Videoradiography
Published on: March 12, 2021
Trapeziometacarpal joint implants can be evaluated by roentgen stereophotogrammetric analysis
T B Hansen1, K Larsen, L Bjergelund
1Department of Orthopaedics, Section of Hand Surgery, Regional Hospital Holstebro, Holstebro, Denmark. tbhansen@dadlnet.dk
Abstract:
Both marker-based roentgen stereophotogrammetric analysis (RSA) and model-based RSA have been helpful evaluation tools in hip and knee arthroplasty. The purpose of this study was to test both model-based and marker-based RSA in the evaluation of total joint prostheses of the trapeziometacarpal joint. In a phantom study, the precision of marker-based RSA was tested with a cemented polyethylene cup and compared with the precision of model-based RSA in an uncemented Elektra screw cup. The precision of model-based RSA of the metacarpal stem was tested using an uncemented Elektra metacarpal stem. In a clinical study 11 patients had double stereo radiographs followed by RSA analysis. The precision of translation in both marker-based and model-based RSA was sufficient for clinical use, but rotation cannot be estimated with sufficient precision.
