Single Arc Volumetric Modulated Arc Therapy of head and neck cancer
Anders Bertelsen1, Christian R Hansen, Jørgen Johansen
1Institute of Clinical Research, University of Southern Denmark, Odense, Denmark. anders.bertelsen@ouh.regionsyddanmark.dk
Background:
The quality of Volumetric Modulated Arc Therapy (VMAT) plans is highly dependent on the performance of the optimization algorithm used. Recently new algorithms have become available which are capable of generating VMAT plans for Elekta accelerators. The VMAT algorithm in Pinnacle is named SmartArc and its capability to generate treatment plans for head and neck cancer was tested.
Methods:
Twenty-five patients with oropharyngeal or hypopharyngeal carcinoma, previously treated with IMRT by means of Pinnacle and Elekta accelerators, were replanned with single arc VMAT. The VMAT planning objectives were to achieve clinical target coverage and sparing of the organs at risk (OAR). Comparison with the original clinically used IMRT was made by evaluating (1) dose-volume histograms (DVHs) for PTVs, (2) DVHs for OARs, (3) delivery time and monitor units (MU), and (4) treatment accuracy.
Results:
Equivalent or superior target coverage and sparing of OARs were achieved with VMAT compared to IMRT. Volumes in the healthy tissues receiving between 17.3 Gy and 49.4 Gy were significantly reduced and the conformity (CI(95%)) of the elective PTV was improved from 1.7 with IMRT to 1.6 with VMAT. Compared to step-and-shoot IMRT, VMAT reduced the number of MUs by 8.5% to 460+/-63 MUs per fraction, and delivered on an Elekta Synergy accelerator, the treatment time was on average reduced by 35% to 241 +/- 16s. In Delta4 measurements of the VMAT treatments, 99.6 +/- 0.5% of the detector points passed a 3 mm and 3% gamma criterion, identical to the results of IMRT.
Conclusions:
The target coverages obtained in the IMRT and VMAT plans were found to be very similar. SmartArc generated single arc VMAT plans with equivalent or better target coverage and sparing of OARs compared to IMRT, while both delivery time and MUs were decreased. Very good dose accuracy results were obtained delivering the plans on an Elekta accelerator.
Related Concept Videos
Arteries of the Head and Neck
The internal carotid arteries supply blood to the anterior portion of the cerebrum. They enter the...
Veins of Head and Neck
On the other hand, the vertebral veins, unlike their arterial counterparts, are not primarily responsible for brain drainage. Instead, they drain the cervical vertebrae, spinal cord, and some small...
Treatment Resistant Cancers
Radiological Investigation II: MRI and Ventilation Perfusion Scan
Magnetic Resonance Imaging (MRI) and Ventilation Perfusion Scans are two radiological investigations that offer detailed diagnostic images of the body, particularly lung structures.
MRI
MRI uses magnetic fields and radiofrequency signals to distinguish between normal and abnormal tissues. This technology provides a more detailed diagnostic image than CT scans, enabling it to characterize pulmonary nodules, stage bronchogenic carcinoma, and evaluate inflammatory activity in...