Jove
Visualize
Contact Us
JoVE
x logofacebook logolinkedin logoyoutube logo
ABOUT JoVE
OverviewLeadershipBlogJoVE Help Center
AUTHORS
Publishing ProcessEditorial BoardScope & PoliciesPeer ReviewFAQSubmit
LIBRARIANS
TestimonialsSubscriptionsAccessResourcesLibrary Advisory BoardFAQ
RESEARCH
JoVE JournalMethods CollectionsJoVE Encyclopedia of ExperimentsArchive
EDUCATION
JoVE CoreJoVE BusinessJoVE Science EducationJoVE Lab ManualFaculty Resource CenterFaculty Site
Terms & Conditions of Use
Privacy Policy
Policies

Related Concept Videos

Controlled-Current Coulometry: Overview01:27

Controlled-Current Coulometry: Overview

Controlled current coulometry, also known as amperostatic coulometry, is a technique used in electrochemical analysis to measure the quantity of a substance through the controlled passage of current. It involves the application of a constant current to an electrochemical cell containing the analyte of interest. As the current flows through the cell, the analyte undergoes a redox reaction at the electrode surface, resulting in a charge transfer. By monitoring the time required for a certain...
Amperometry: Overview01:10

Amperometry: Overview

Amperometry is a technique commonly used to measure the concentration of specific analytes in a solution by monitoring the electric current generated during an electrochemical reaction. It involves applying a constant potential between a working electrode and a reference electrode to measure the resulting current, which is proportional to the concentration of the analyte. The Clark oxygen electrode operates based on this principle of amperometry. It consists of a cathode and an anode enclosed...
High-Performance Liquid Chromatography: Types of Detectors01:15

High-Performance Liquid Chromatography: Types of Detectors

The role of the detectors in High-Performance Liquid Chromatography (HPLC) is to analyze the solutes as they exit from the chromatographic column. The detector recognizes the solute's property and generates corresponding electrical signals, which are converted into a readable graph of the detector's response versus elution time called a chromatogram at the computer. There are several types of HPLC detectors, each with its own advantages and limitations, depending on the analyte properties and...
Precipitation Titration: Endpoint Detection Methods01:19

Precipitation Titration: Endpoint Detection Methods

In argentometric precipitation titrations, endpoints can be detected visually by the Mohr, Volhard, and Fajans methods. In the Mohr method, adding a soluble chromate indicator gives an initial yellow color to the analyte solution. As the titrant is added, the first excess of silver ions forms a red silver chromate precipitate, marking the endpoint. The solution pH should be maintained at about 8 by adding solid CaCO3.
In the Volhard method, a standard excess of AgNO3 is first added to the...
Gas Chromatography: Types of Detectors-I01:21

Gas Chromatography: Types of Detectors-I

There are different types of detectors used in gas chromatography, each with its own specific properties that make it suitable for detecting certain types of analytes. The most commonly used detectors in GC are thermal conductivity detector (TCD), flame ionization detector (FID), and electron capture detector (ECD).
TCD is the earliest and most widely used detector that operates by measuring the changes in the thermal conductivity of the carrier gas. When a sample compound enters the detector,...
Flame Photometry: Lab01:16

Flame Photometry: Lab

In a flame photometer, when a solution like potassium chloride is aspirated into the flame, the solvent evaporates, leaving behind dehydrated salt. This salt dissociates into free gaseous atoms in their ground state. Some of these atoms absorb energy from the flame, leading to their excitation. The excited atoms return to the ground state, emitting photons at characteristic wavelengths. Because only electronic transitions are involved, the resulting emission lines are very narrow. The intensity...

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Advanced Separators for Liquid and Quasi-Solid Lithium Rechargeable Batteries: Design and Development.

Advanced materials (Deerfield Beach, Fla.)·2026
Same author

A voltammetric method for determining N-nitrosodimethylamine in losartan and olmesartan drugs.

Journal of pharmaceutical sciences·2025
Same author

One-step microwave-assisted synthesis of fluorescent carbon quantum dots for determination of ascorbic acid and riboflavin in vitamin supplements.

Spectrochimica acta. Part A, Molecular and biomolecular spectroscopy·2024
Same author

The effects of heat stress on intrauterine development, reproductive function, and ovarian gene expression of F1 female mice as well as gene expression of F2 embryos†.

Biology of reproduction·2023
Same author

Fast automated method for the direct determination of total antimony in grape juice samples by hydride generation and atomic fluorescence spectrometric detection without external pretreatment.

Food chemistry·2022
Same author

Simultaneous determination of methyl, ethyl, propyl, and butyl parabens in sweetener samples without any previous pretreatment using square wave voltammetry and multiway calibration.

Food chemistry·2021

Related Experiment Video

Updated: Jun 15, 2026

Dynamic Electrochemical Measurement of Chloride Ions
07:32

Dynamic Electrochemical Measurement of Chloride Ions

Published on: February 5, 2016

Automatic determination of chlorine without standard solutions using a biamperometric flow-batch analysis system.

Valberes B Nascimento1, Thiago M G Selva, Elaine C S Coelho

  • 1Universidade Federal Rural de Pernambuco, Departamento de Química, 52171-030, Recife, PE, Brazil. valberes@dq.ufrpe.br

Talanta
|March 2, 2010
PubMed
Summary

An automated system generates standards in-line for chlorine analysis, simplifying measurements in real samples. This accurate, cost-effective method offers reliable results for routine applications.

More Related Videos

Automated, High-resolution Mobile Collection System for the Nitrogen Isotopic Analysis of NOx
07:14

Automated, High-resolution Mobile Collection System for the Nitrogen Isotopic Analysis of NOx

Published on: December 20, 2016

Dry Film Photoresist-based Electrochemical Microfluidic Biosensor Platform: Device Fabrication, On-chip Assay Preparation, and System Operation
13:42

Dry Film Photoresist-based Electrochemical Microfluidic Biosensor Platform: Device Fabrication, On-chip Assay Preparation, and System Operation

Published on: September 19, 2017

Related Experiment Videos

Last Updated: Jun 15, 2026

Dynamic Electrochemical Measurement of Chloride Ions
07:32

Dynamic Electrochemical Measurement of Chloride Ions

Published on: February 5, 2016

Automated, High-resolution Mobile Collection System for the Nitrogen Isotopic Analysis of NOx
07:14

Automated, High-resolution Mobile Collection System for the Nitrogen Isotopic Analysis of NOx

Published on: December 20, 2016

Dry Film Photoresist-based Electrochemical Microfluidic Biosensor Platform: Device Fabrication, On-chip Assay Preparation, and System Operation
13:42

Dry Film Photoresist-based Electrochemical Microfluidic Biosensor Platform: Device Fabrication, On-chip Assay Preparation, and System Operation

Published on: September 19, 2017

Area of Science:

  • Analytical Chemistry
  • Electrochemistry

Background:

  • Traditional methods for chlorine analysis often require standard solutions.
  • Developing automated, reliable, and cost-effective analytical systems is crucial for routine applications.

Purpose of the Study:

  • To present an automatic analysis system for determining total available chlorine.
  • To demonstrate the system's versatility and applicability to real-world samples.

Main Methods:

  • Utilized an electrochemical flow cell for in-line standard generation.
  • Employed the standard addition technique with biamperometric monitoring.
  • Optimized flow-batch parameters for sensitivity and precision.

Main Results:

  • Achieved excellent linear dependence between biamperometric signal and chlorine concentration.
  • Demonstrated good agreement with a reference method (r=0.9995).
  • Obtained a low limit of detection (LOD=8.261 x 10(-7) mol L(-1)) and recoveries near 100%.

Conclusions:

  • The developed system provides a simple, accurate, and inexpensive method for chlorine determination.
  • The system is highly reliable and suitable for routine applications.
  • The method can be extended to other oxidants and sample types.