A head phantom study for intraocular dose evaluation of 64-slice multidetector CT examination in patients with
Kosuke Matsubara1, Kichiro Koshida, Kimiya Noto
1Department of Quantum Medical Technology, Faculty of Health Sciences, Kanazawa University, 5-11-80 Kodatsuno, Kanazawa, Ishikawa 920-0942, Japan. matsuk@mhs.mp.kanazawa-u.ac.jp
Purpose:
In cases of suspected cranial trauma, cranial CT examinations should be performed to rule out pathology. There are some methods available for reducing intraocular doses; however, it is difficult for the operators to conduct the necessary measurements because of restrictions in time and patient mobility, especially in high-energy trauma cases. Therefore, we performed a head phantom study for intraocular dose evaluation of 64-slice multidetector CT examination in patients with suspected cranial trauma.
Materials And Methods:
Assuming that the orbitomeatal (OM) line and bed were vertical, a head phantom was tilted from 10 degrees caudally to 25 degrees cranially at 5-degree intervals. At each tilted position, the phantom was examined using a 64-section multidetector CT device using three acquisition protocols. Intraocular doses during each examination were measured using small dosimeters.
Results:
Assuming that the OM line and bed were vertical, intraocular doses varied between 52 and 140%, 17-138%, and 90-142% during helical, non-helical, and helical CT angiographic examinations, respectively. Intraocular doses increased when the phantom was tilted cranially.
Conclusion:
If possible, the best way to reduce the intraocular dose is by angling the gantry cranially, tilting the head of each patient caudally and adopting a non-helical acquisition method. During procedure, the acquisition angle should be angled cranially more than 0 degrees based on the OM line. The estimation of intraocular dose using the acquisition angle and displayed volumetric CT dose index might be useful to evaluate the deterministic effect risks and to inform patients about the associated risks.
More Related Videos
Related Concept Videos
Imaging Studies III: Computed Tomography
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies for Cardiovascular System V: CT
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...