Related Experiment Video
Updated: Jun 15, 2026

Methylated DNA Immunoprecipitation
Published on: January 2, 2009
A hydrophilic interaction ultraperformance liquid chromatography (HILIC-UPLC) method for genomic DNA methylation
Salvatore Sotgia1, Angelo Zinellu, Elisabetta Pisanu
1Department of Biomedical Sciences and Centre of Excellence for Biotechnology Development and Biodiversity Research, University of Sassari, 07100 Sassari, Italy. salvatore.sotgia@gmail.com
Abstract:
A hydrophilic interaction chromatography-based method, in combination with 1.7 microm ethylene bridged hybrid particle packed column (100 mm x 2.1 mm I.D.) and ultraperformance liquid chromatography, has been developed to measure cytosine (C) and methylcytosine (mC) in order to evaluate the extent of DNA methylation. Separation of cytosine and methylcytosine was achieved with good resolution and in fairly short times (5.5 min) by using isocratic elution with a mixture of 97:3 (v/v) acetonitrile/10 mM ammonium acetate as a mobile phase. The determination coefficients of C and mC were high (R(2) > 0.999) within the range tested. The %RSD for intraday and interday were respectively 2.2% and 2.5% for C and 3.5% and 3.8% for mC. The limit of detection was 0.52 microM (0.52 fmol on-column) both for C and mC while the limit of quantification was 1.72 microM (1.72 fmol on-column) both for C and mC. The smallest amount of purified DNA that yielded a measurable level of C and mC was 10 microg. On the whole, this method is simple, rapid, sensitive, and precise.