In-vitro assessment of temperature rise in the pulp during orthodontic bonding
Siddik Malkoç1, Tancan Uysal, Serdar Uşümez
1Department of Orthodontics, and Chair, School of Dentistry, University of Inönü, Malatya, Turkey. siddikmalkoc@yahoo.com
Introduction:
In this in-vitro study, we evaluated the temperature changes in the pulp chamber during bracket bonding using 4 different light sources.
Methods:
Eighty intact extracted maxillary central incisors were used. The teeth were divided into 4 groups of 20 teeth each. Brackets (Mini Twin, Dentaurum, Ispringen, Germany) were bonded with Transbond XT (3M Unitek, Monrovia, Calif) adhesive and light cured with low-intensity halogen light for 40 seconds, high-intensity halogen light for 40 seconds, light-emitting diode (LED) light for 20 seconds, and plasma arc light (PAC) for 6 seconds. Light curing was performed 5 mm from tooth surfaces. A J-type thermocouple wire was positioned in the center of the pulp chamber. The results were analyzed with analysis of variance (ANOVA) and the Tukey HSD test.
Results:
ANOVA and the Tukey HSD test showed that pulp chamber temperature changes were influenced by the type of light source. All groups showed significant differences between each other (P <0.001). The intrapulpal temperature changes induced by different light sources were the following: high-intensity halogen (6.84 degrees C +/- 2.44 degrees C), low-intensity halogen (4.71 degrees C +/- 0.96 degrees C), LED (2.95 degrees C +/- 1.12 degrees C), and PAC (0.96 degrees C +/- 0.83 degrees C).
Conclusions:
High- and low-intensity halogen light induced significantly higher intrapulpal temperature changes than did the LED and PAC. Except for the high intensity halogen light, orthodontic bonding with light-curing units did not exceed the critical 5.5 degrees C rise in temperature reported to produce pulpal damage.
Related Concept Videos
Assessing Body Temperature - Oral
Step 1:
Start by practicing proper hand hygiene to prevent the spread of microorganisms.
Step 2:
Take the thermometer out of the charging unit, switch it on, and wait for the ready sign.
Step 3:
Gently slide the probe cover until a click is heard. This simple action prevents cross-contamination and ensures the correct placement of the probe cover.
Step 4:
Instruct the patient to open their mouth and place...
Assessing Body Temperature - Temporal Artery
Step 1: Perform hand hygiene and don a fresh pair of gloves to prevent cross-infection and ensure patient safety.
Step 2: Explain the procedure to the patient to establish trust. Clear communication establishes trust with the patient, ensures they understand what to expect, promotes cooperation, and enhances comfort during the procedure.
Step 3: Assess the patient's forehead...