Related Experiment Video
Updated: Jun 15, 2026

04:41
Treatment with Locking Intramedullary Nailing for Intertrochanteric Fracture of the Femur Utilizing a New Awl with a Distal Positioner
Published on: June 6, 2025
Early experience with a novel nonmetallic cable in reconstructive hip surgery
Nicholas T Ting1, Glenn D Wera, Brett R Levine
1Department of Orthopaedic Surgery, Rush University Medical Center, 1725 West Harrison, Suite 1063, Chicago, IL 60612, USA.
Clinical Orthopaedics and Related Research
|March 6, 2010
Summary
This study found that novel nonmetallic cerclage cables provided adequate fixation for periprosthetic fractures and osteotomies during total hip arthroplasty (THA), with no cable-related complications observed.
Area of Science:
- Orthopedic Surgery
- Biomaterials Science
- Medical Device Innovation
Background:
- Metallic wires and cables are standard for THA fixation but pose risks like fraying and wear.
- Concerns include third-body generation, accelerated bearing surface wear, and surgical team injury.
- Novel nonmetallic cerclage cables offer a potential alternative to address these issues.
Purpose of the Study:
- To evaluate the efficacy of nonmetallic cerclage cables in THA.
- To determine the rates of cable failure, union, and complications.
- To assess the use of these cables in periprosthetic fracture and osteotomy fixation.
Main Methods:
- Retrospective review of 29 patients undergoing primary or revision THA with nonmetallic cables.
- Indications included extended trochanteric osteotomy, femur fractures, and strut allograft fixation.
- Clinical and radiographic follow-up for a minimum of 13 months.
Main Results:
- Two of 29 patients (7%) experienced nonunion; other fractures/osteotomies healed.
- Four patients (14%) had postoperative dislocations, two requiring reoperation.
- No cable breakage or complications directly linked to the nonmetallic cables were identified.
Conclusions:
- Nonmetallic periprosthetic cables offer adequate fixation for osteotomy and fracture healing in THA.
- No complications were directly attributed to the use of these novel cables.
- This suggests a promising alternative to traditional metallic fixation systems.