Related Experiment Video
Updated: Jun 15, 2026

Polytetrafluoroethylene (PTFE) as a Suture Material in Tendon Surgery
Published on: October 6, 2022
Mechanical strength of the side-to-side versus Pulvertaft weave tendon repair
Stephen H M Brown1, Eric R Hentzen, Alan Kwan
1Departments of Orthopaedic Surgery and Radiology, University of California, San Diego, San Diego, CA, USA.
Purpose:
The side-to-side (SS) tendon suture technique was designed to function as a repair that permits immediate postoperative activation and mobilization of a transferred muscle. This study was designed to test the strength and stiffness of the SS technique against a variation of the Pulvertaft (PT) repair technique.
Methods:
Flexor digitorum superficialis (FDS) and flexor digitorum profundus (FDP) tendons were harvested from 4 fresh cadavers and used as a model system. Seven SS and 6 PT repairs were performed, using the FDS as the donor and the FDP as the recipient tendon. For SS repairs, the FDS was woven through one incision in the FDP and was joined with 4 cross-stitch running sutures down both sides and one double-loop suture at each tendon free end. For PT repairs, the FDS was woven through 3 incisions in the FDP and joined with a double-loop suture at both ends of the overlap and 4 evenly spaced mattress sutures between the ends. Tendon repairs were placed in a tensile testing machine, preconditioned, and tested to failure.
Results:
There were no statistically significant differences in cross-sectional area (p = .99) or initial length (p = .93) between SS and PT repairs. Therefore, all comparisons between methods were made using measures of loads and deformations, rather than stresses and strains. All failures occurred in the repair region, rather than at the clamps. However, failure mechanisms were different between the 2 techniques-PT repairs failed by the suture knots either slipping or pulling through the tendon material, followed by the FDS tendon pulling through the FDP tendon; SS repairs failed by shearing of fibers within the FDS. Load at first failure, ultimate load, and repair stiffness were all significantly different between SS and PT techniques; in all cases, the mean value for SS was higher than for PT.
Conclusions:
The SS repair using a cross-stitch suture technique was significantly stronger and stiffer than the PT repair using a mattress suture technique. This suggests that using SS repairs could enable patients to load the repair soon after surgery. Ultimately, this should reduce the risk of developing adhesions and result in improved functional outcome and fewer complications in the acute postoperative period. Future work will address the specific mechanisms (eg, suture-throw technique and tendon-weave technique) that underlie the improved strength and stiffness of the SS repair.
Related Concept Videos
Tensile Strength Considerations of Concrete
The dimensions and shape of a concrete specimen also...
Relation Between Tensile Strength and Compressive Strength of Concrete
Posttensioned Masonry Walls
Post-tensioned masonry walls use high-strength steel rods or flexible tendons to enhance the strength and efficiency of masonry structures. These elements are securely anchored to the foundation and extend vertically either within the cores of the masonry units or between the masonry wythes. The construction process involves building the wall with these tensioning elements in place and allowing the mortar to fully cure.
Following the curing process, the tensioning begins. Steel rods are...
Fractures: Bone Repair
Minor fractures with no bone displacement are treated by immobilizing the fractured bone using a cast or splint. However, in the case of fractures with displaced bones, the broken bones are repositioned before immobilization to ensure successful healing without deformation and loss of function. The realignment of fractured bone ends is performed through a process called reduction. If the procedure...