Related Experiment Video
Updated: Jun 15, 2026

05:46
Introduction of Intracapsular Rotary-cut Procedures (IRCP): A Modified Hysteromyomectomy Procedures Facilitating Fertility Preservation
Published on: January 17, 2019
Randomized controlled study comparing the Gynecare Morcellex and Rotocut G1 tissue morcellators
Fulvio Zullo1, Angela Falbo, Assunta Iuliano
1Department of Obstetrics, University Magna Graecia of Catanzaro, Catanzaro, Italy.
Journal of Minimally Invasive Gynecology
|March 16, 2010
Summary
The Gynecare Morcellex tissue morcellator demonstrated effectiveness in laparoscopic hysterectomy and myomectomy procedures. It offers excellent safety and handling compared to the Rotocut G1 morcellator.
Area of Science:
- Gynecologic surgery
- Minimally invasive procedures
Background:
- Laparoscopic supracervical hysterectomy and myomectomy are common procedures for symptomatic uterine fibroids.
- Efficient tissue morcellation is crucial for successful outcomes in these surgeries.
Purpose of the Study:
- To evaluate the effectiveness, safety, and handling of the Gynecare Morcellex tissue morcellator.
- To compare the Gynecare Morcellex with the Rotocut G1 morcellator in laparoscopic procedures.
Main Methods:
- A randomized controlled trial was conducted with 74 patients undergoing laparoscopic supracervical hysterectomy or myomectomy.
- Patients were divided into two groups: one using the Gynecare Morcellex (experimental) and the other using the Rotocut G1 (control).
- Surgical data and handling scores were assessed for both groups.
Main Results:
- No significant differences were observed in patient characteristics or surgical data between the Gynecare Morcellex and Rotocut G1 groups.
- The Gynecare Morcellex group showed a significantly better handling score (p < .05) compared to the control group.
Conclusions:
- The Gynecare Morcellex is an effective tissue morcellator for laparoscopic hysterectomy and myomectomy.
- The instrument offers excellent safety and superior handling characteristics.