Related Concept Videos

Computed Tomography01:10

Computed Tomography

Tomography refers to imaging by sections. Computed tomography (CT) is a non-invasive imaging technique that uses computers to analyze several cross-sectional X-rays to reveal minute details about structures in the body.
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Matrixed-Spectrum Decomposition Accelerated Linear Boltzmann Transport Equation Solver for Fast Scatter Correction in Multi-Spectral CT.

IEEE transactions on medical imaging·2026
Same author

A comparative analysis of two dual-layer flat panel detector designs and mimicking of a triple-layer flat panel detector in spectral CT.

Medical physics·2026
Same author

Energy-Threshold Bias Calculator: A Physics-Model-Based Adaptive Correction Scheme for Photon-Counting CT.

IEEE transactions on medical imaging·2026
Same author

Disentangled deep learning method for interior tomographic reconstruction of low-dose x-ray CT.

Physics in medicine and biology·2025
Same author

Improved fast kV-switching dual-energy CBCT with a spectral modulator: System design and a feasibility study.

Medical physics·2025
Same author

ComptoNet: a Compton-map guided deep learning framework for multi-scatter estimation in multi-source stationary CT.

Physics in medicine and biology·2025
Same journal

Correction to "On the shape of the radiation survival curve in tumor spheroids: The role of oxygen heterogeneity".

Medical physics·2026
Same journal

Multi-view constrained semi-supervised vertebra detection for 3D ultrasound spine volume.

Medical physics·2026
Same journal

Accuracy of quantitative <sup>177</sup>Lu SPECT/CT imaging: A systematic review.

Medical physics·2026
Same journal

Physics-constrained dual-domain network for CBCT reconstruction from orthogonal X-rays in gynecologic radiotherapy.

Medical physics·2026
Same journal

Decomposition-based harmonization for quantitative PET imaging across scanners and radiotracers.

Medical physics·2026
Same journal

Development and evaluation of an in vivo dose-based monitoring system for electron FLASH radiation therapy.

Medical physics·2026
See all related articles

Related Experiment Video

Updated: Jun 15, 2026

X-ray Dose Reduction through Adaptive Exposure in Fluoroscopic Imaging
08:30

X-ray Dose Reduction through Adaptive Exposure in Fluoroscopic Imaging

Published on: September 11, 2011

Scatter correction method for x-ray CT using primary modulation: phantom studies.

Hewei Gao1, Rebecca Fahrig, N Robert Bennett

  • 1Department of Radiology, Stanford University, Stanford, California 94305, USA. heweigao@stanford.edu

Medical Physics
|March 17, 2010
PubMed
Summary
This summary is machine-generated.

This study demonstrates a primary modulation technique for scatter correction in X-ray computed tomography (CT). The method significantly reduces CT number errors and improves contrast-to-noise ratio in challenging imaging scenarios.

More Related Videos

Multimodal Cross-Device and Marker-Free Co-Registration of Preclinical Imaging Modalities
07:13

Multimodal Cross-Device and Marker-Free Co-Registration of Preclinical Imaging Modalities

Published on: October 27, 2023

Related Experiment Videos

Last Updated: Jun 15, 2026

X-ray Dose Reduction through Adaptive Exposure in Fluoroscopic Imaging
08:30

X-ray Dose Reduction through Adaptive Exposure in Fluoroscopic Imaging

Published on: September 11, 2011

Multimodal Cross-Device and Marker-Free Co-Registration of Preclinical Imaging Modalities
07:13

Multimodal Cross-Device and Marker-Free Co-Registration of Preclinical Imaging Modalities

Published on: October 27, 2023

Area of Science:

  • Medical Imaging
  • Radiological Physics
  • X-ray Computed Tomography

Background:

  • Scatter correction is a critical challenge in X-ray imaging, particularly with large area detectors.
  • Previous work introduced a primary modulation method for scatter correction in CT, showing initial promise.
  • This study aims to quantitatively evaluate the primary modulation technique in more complex imaging scenarios.

Purpose of the Study:

  • To quantitatively evaluate the performance of a primary modulation scatter correction method in X-ray CT.
  • To assess the method's effectiveness in challenging applications with significant scatter.
  • To analyze potential errors in scatter estimation within the primary modulation technique.

Main Methods:

  • Investigated the primary modulation method on two tabletop CT systems using three phantoms (Catphan 600, anthropomorphic chest, and modified Catphan 600).
  • Analyzed potential errors in scatter estimation associated with the primary modulation technique.
  • Employed two primary modulators (aluminum and copper) with varying modulation strengths and frequencies to assess their impact on scatter correction efficiency.

Main Results:

  • Reduced CT number errors from 371.4 to 21.9 HU and increased contrast-to-noise ratio from 10.9 to 19.2 on a Catphan 600 phantom.
  • Significantly reduced CT number errors on an anthropomorphic chest phantom (from 327 to 19 HU) in a challenging high-scatter scenario.
  • Demonstrated that scatter-to-primary ratio estimation error increases with object size, rising from 0.04 to 0.1 with larger phantoms.

Conclusions:

  • The primary modulation method shows good scatter correction performance across various phantom studies, confirmed by quantitative analysis and image comparisons.
  • A strong primary modulation, characterized by a low transmission factor and high modulation frequency, is optimal for achieving high scatter correction accuracy.
  • The technique is effective in reducing scatter-induced artifacts in X-ray CT, even in complex and heterogeneous objects.