Dose optimisation of double-contrast barium enema examinations
1Department of Medical Physics and Biomedical Engineering, Sahlgrenska University Hospital, Gothenburg, Sweden. berner_karin@hotmail.com
Abstract:
The purpose of the present work was to optimise the filtration and dose setting for double-contrast barium enema examinations using a Philips MultiDiagnost Eleva FD system. A phantom study was performed prior to a patient study. A CDRAD phantom was used in a study where copper and aluminium filtration, different detector doses and tube potentials were examined. The image quality was evaluated using the software CDRAD Analyser and the phantom dose was determined using the Monte Carlo-based software PCXMC. The original setting [100 % detector dose (660 nGy air kerma) and a total filtration of 3.5 mm Al, at 81 kVp] and two other settings identified by the phantom study (100 % detector dose and additional filtration of 1 mm Al and 0.2 mm Cu as well as 80 % detector dose and added filtration of 1 mm Al and 0.2 mm Cu) were included in the patient study. The patient study included 60 patients and up to 8 images from each patient. Six radiologists performed a visual grading characteristics study to evaluate the image quality. A four-step scale was used to judge the fulfillment of three image quality criteria. No overall statistical significant difference in image quality was found between the three settings (P > 0.05). The decrease in the effective dose for the settings in the patient study was 15 % when filtration was added and 34 % when both filtrations was added and detector dose was reduced. The study indicates that additional filtration of 1 mm Al and 0.2 mm Cu and a decrease in detector dose by 20 % from the original setting can be used in colon examinations with Philips MultiDiagnost Eleva FD to reduce the patient dose by 30 % without significantly affecting the image quality. For 20 exposures, this corresponds to a decrease in the effective dose from 1.6 to 1.1 mSv.
Related Concept Videos
Lower GI Series: Barium Enema
Procedure Details
The examination begins by inserting a lubricated rectal tube into the patient's rectum to administer a radiopaque barium solution. The barium flow is carefully...
Upper GI Series: Barium Swallow
Purpose and Procedure
Patients undergoing this procedure ingest a liquid containing barium sulfate with a chalky...
Imaging Studies III: Computed Tomography
Imaging Studies III: Gastrointestinal Motility Studies and Virtual Colonoscopy
Radionuclide Testing
Radionuclide testing is a sophisticated medical technique for assessing gastrointestinal motility. It focuses on gastric emptying and colonic transit time. Radioactive markers track the movement of food through the digestive system, providing insights into gastrointestinal disorders.
In gastric emptying studies, a meal's liquid and solid...
Radiological Investigation I: X-ray and CT
Imaging Studies for Cardiovascular System V: CT