Related Experiment Video
Updated: Jun 14, 2026

A Methodological Protocol and Considerations for Transcranial Ultrasonic Stimulation in Exploratory Clinical Human Studies
Published on: December 12, 2025
Initial clinical experience of an ultrasonic strain imaging system with novel noise-masking capability
1Department of Engineering, University of Cambridge, Trumpington Street, Cambridge CB2 1PZ, UK.
Abstract:
Quasistatic strain imaging is a form of elastography that can produce qualitative images of tissue stiffness with only software modifications to conventional ultrasound hardware. Unlike current commercial offerings, the novel strain-imaging system that is the subject of this paper displays regions of signal decorrelation using an overlaid colour mask and can also produce three-dimensional (3D) strain images. In illustrative studies of the breast, testis and thyroid, the colour mask is seen to reduce the potential to misinterpret noise as meaningful stiffness information, and also helps to differentiate cystic and solid lesions. High-quality imaging of the testis in vivo demonstrates that 3D strain imaging is feasible.
More Related Videos
16:01An Experimental Protocol for Assessing the Performance of New Ultrasound Probes Based on CMUT Technology in Application to Brain Imaging
Published on: September 24, 2017
07:52Pipeline for Planning and Execution of Transcranial Ultrasound Neuromodulation Experiments in Humans
Published on: June 28, 2024