Related Experiment Video
Updated: Jun 14, 2026

A Rapid and Chemical-free Hemoglobin Assay with Photothermal Angular Light Scattering
Published on: December 7, 2016
Detection of Hb Constant Spring by a capillary electrophoresis method
Can Liao1, Jian-Ying Zhou, Xing-Mei Xie
1Prenatal Diagnostic Center, Guangzhou Maternal & Neonatal Hospital, Guangzhou Women & Children Medical Center Guangzhou Medical College, Guangzhou, Guangdong, People's Republic of China.
Abstract:
Hb Constant Spring [Hb CS; alpha142, Term-->Gln (TAA>CAA in alpha2)] is the most prevalent nondeletional alpha-thalassemia (alpha-thal) anomaly in southern China. In conjunction with alpha(0)-thal, it can cause severe Hb H (beta(4)) disease. The present study was done to evaluate the efficiency of two diagnostic methods in detecting Hb CS. Automated high performance liquid chromatography (HPLC) and Sebia Capillarys 2, a capillary electrophoresis method, were applied to blood samples from 21 individuals with Hb CS trait. Of the 21 cases, all (100%) were detected by capillary electrophoresis, whereas only 16 (76.2%) were detected by HPLC. We concluded that the Sebia Capillarys 2 is the preferred method for Hb CS screening.