Validation of the central blood pressure estimation by the SphygmoCor system in Chinese
Jun-Li Zuo1, Yan Li, Zi-Jun Yan
1Department of Hypertension, Ruijin Hospital, Shanghai Jiaotong University School of Medicine, Shanghai, China.
Background:
Pulse wave analysis using the SphygmoCor system allows the estimation of central blood pressures (BP). However, there is controversy over its accuracy for clinical use.
Methods:
In 45 patients undergoing coronary angiography, we compared the ascending aorta BPs measured by the invasive catheter with the estimations by the SphygmoCor system, using paired t-tests, simple correlation analysis, and Bland-Altman plots.
Results:
The estimation of central systolic BP by SphygmoCor was lower, although statistically insignificant, than that measured by the catheter (144±29 vs. 148±30 mmHg; P=0.98). The standard deviation of the difference amounted to 17 mmHg. Both the measured and estimated central systolic BPs were significantly (P<0.001) lower than the brachial systolic BP (156±30 mmHg). The diastolic BP measured at the brachial artery (87±15 mmHg) and provided by SphygmoCor (90±15 mmHg) were systematically higher (P<0.001) than that measured by the catheter (74±13 mmHg). The SphygmoCor system underestimated the central pulse pressure by -20±14 mmHg compared with the catheter method. The correlation coefficients for systolic BP, diastolic BP, and pulse pressure between catheter measurements and the SphygmoCor estimations were 0.84, 0.60, and 0.82 (P<0.001), respectively.
Conclusion:
When the radial waveform was calibrated with the oscillometric brachial pressures, the SphygmoCor system could not provide accurate estimation of central BPs. The inaccurate measurement of cuff pressure was the major limiting factor for the use of the transfer function in the clinical settings.
Related Concept Videos
Assessment of blood pressure in brachial artery(two-step method)
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Measurement of Blood Pressure
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement