Related Experiment Video
Updated: Jun 14, 2026

Synthesis of Programmable Main-chain Liquid-crystalline Elastomers Using a Two-stage Thiol-acrylate Reaction
Published on: January 19, 2016
Peak parking determination of the obstruction factor in lauryl acrylate monolithic CEC columns
Gwendolyn J Anderson1, Zoe LaPier, Michael B Cammarata
1Department of Chemistry, Trinity University, San Antonio, TX 78212, USA.
Abstract:
The peak parking method was used to determine the obstruction factor of lauryl acrylate porous polymer monoliths. Polymers were prepared in situ in fused-silica capillaries using thermally initiated polymerization. These columns have been used for CEC of neutral analytes. Thiourea, which is unretained, was used as the test analyte for the obstruction factor measurement. The obstruction factor was determined to be 0.72 with a SD of (+/-0.01), which is consistent with the concept that organic porous polymer monoliths are more permeable than traditional LC stationary phases.
Related Concept Videos
Gas Chromatography: Types of Columns and Stationary Phases
For an analyte to remain on the column for a sufficient amount of time, it must exhibit some level of compatibility (or...
Silica Gel Column Chromatography: Overview