A new ground and reference technique for invasive EEG recordings
Serge Vulliemoz1, Laurent Spinelli, Daniel Pellise
1Presurgical Epilepsy Evaluation Unit, Department of Neurology, University Hospital of Geneva, Geneva, Switzerland.
Objective:
To demonstrate a subdermal wire electrode technique for establishing a ground (GND) and reference (REF) during long-term EEG monitoring (LTM) with intracranial electrodes. Usually, a separate GND and REF are required and this GND&REF pair can be selected contacts in the invasive electrode arrays themselves, special invasive electrodes, or standard surface disc electrodes which require frequent maintenance. We investigated the use of a pair of chronic Subdermal Wire Electrodes (SWE) for use as GND&REF.
Methods:
A pair of SWEs as GND&REF was tested in nine patients undergoing invasive EEG monitoring. SWE impedances were monitored in two patients and compared to disc electrode impedances.
Results:
Without maintenance, SWE impedances remained low and stable during the entire recording period (up to 20 days), whereas disc electrodes showed rapid impedance increase after the first day. In all nine patients, the consistent and durable integrity of the GND&REF pair of SWE allowed for a good quality EEG recording. No local skin complications were observed.
Conclusions:
A pair of SWE electrodes provides a GND&REF system that is easy to place, maintain, and provides a high quality recording with long-term stability when coupled with referential based EEG recording system from invasive electrodes.
Significance:
A more efficient means of establishing a GND&REF pair during the monitoring of patients with invasive electrodes is described.
More Related Videos
06:19Visual Evoked Potential Recordings in Mice Using a Dry Non-invasive Multi-channel Scalp EEG Sensor
Published on: January 12, 2018
08:25Combined Invasive Subcortical and Non-invasive Surface Neurophysiological Recordings for the Assessment of Cognitive and Emotional Functions in Humans
Published on: May 19, 2016
Related Concept Videos
Taping Over Different Ground Profiles
Inertial Frames of Reference
Non-inertial Frames of Reference
Chemical Shift: Internal References and Solvent Effects
The internal reference compound generally used in NMR spectroscopy is tetramethylsilane (TMS). TMS is preferred because it is chemically inert, soluble in NMR solvents, and easily removable. Also, the highly shielded methyl protons in TMS yield an intense...
Connective Tissue Fibers and Ground Substance
Data Reporting and Recording