Noninvasive EEG Recordings from Freely Moving Piglets
Published on: July 13, 2018
A new ground and reference technique for invasive EEG recordings.
Serge Vulliemoz1, Laurent Spinelli, Daniel Pellise
1Presurgical Epilepsy Evaluation Unit, Department of Neurology, University Hospital of Geneva, Geneva, Switzerland.
Chronic subdermal wire electrodes (SWE) offer a stable and low-maintenance ground (GND) and reference (REF) for long-term EEG monitoring. This technique ensures high-quality intracranial EEG recordings without complications.
Area of Science:
- Neuroscience
- Biomedical Engineering
Background:
- Long-term EEG monitoring (LTM) with intracranial electrodes typically requires a separate ground (GND) and reference (REF).
- Traditional GND & REF methods using invasive electrode arrays or surface discs often need frequent maintenance and can suffer from impedance issues.
Purpose of the Study:
- To evaluate the efficacy of chronic subdermal wire electrodes (SWE) as a combined GND and REF system for LTM.
- To demonstrate a novel, low-maintenance technique for establishing stable electrical references during invasive EEG.
Main Methods:
- A prospective study involving nine patients undergoing invasive EEG monitoring.
- A pair of subdermal wire electrodes (SWE) were utilized as the GND & REF.
- Impedance monitoring of SWE was performed in two patients and compared against standard disc electrodes.
Main Results:
- Subdermal wire electrodes (SWE) demonstrated consistently low and stable impedances throughout the recording period (up to 20 days) without maintenance.
- In contrast, disc electrodes exhibited a rapid increase in impedance after the first day.
- All nine patients achieved good quality EEG recordings with the SWE GND & REF system, showing no local skin complications.
Conclusions:
- Chronic subdermal wire electrodes (SWE) provide a simple, maintainable, and stable GND & REF system for LTM.
- This technique ensures high-quality EEG recordings when used with referential-based systems and invasive electrodes.
- The study presents a more efficient method for establishing GND & REF pairs in patients requiring invasive EEG monitoring.
More Related Videos
06:19Visual Evoked Potential Recordings in Mice Using a Dry Non-invasive Multi-channel Scalp EEG Sensor
Published on: January 12, 2018
08:25Combined Invasive Subcortical and Non-invasive Surface Neurophysiological Recordings for the Assessment of Cognitive and Emotional Functions in Humans
Published on: May 19, 2016
Related Concept Videos
Taping Over Different Ground Profiles
Inertial Frames of Reference
Non-inertial Frames of Reference
Chemical Shift: Internal References and Solvent Effects
The internal reference compound generally used in NMR spectroscopy is tetramethylsilane (TMS). TMS is preferred because it is chemically inert, soluble in NMR solvents, and easily removable. Also, the highly shielded methyl protons in TMS yield an intense...
Connective Tissue Fibers and Ground Substance
Data Reporting and Recording