Quantitative Detection of Trace Explosive Vapors by Programmed Temperature Desorption Gas Chromatography-Electron Capture Detector
Published on: July 25, 2014
Parametric evaluation for programmable temperature vaporization large volume injection in gas chromatographic
Hua Wei1, Priyanthi S Dassanayake, An Li
1School of Public Health, University of Illinois at Chicago 2121 West Taylor Street, Chicago, IL 60612.
Abstract:
This work is a thorough investigation on the major operating parameters of the programmable temperature vaporization (PTV) inlet used for gas chromatographic injection, including injection mode and volume, inlet temperature, vent and purge flow rates. The results clearly demonstrate the advantage of large volume injection in enhancing the detection of polybrominated diphenyl ethers (PBDEs). Partial loss of injected PBDEs occurred during solvent venting and due to incomplete sample transfer. Such loss was minimized by lowering the initial inlet temperature and vent flow and elevating the final inlet temperature. The results show that 50 mL/min vent flow, as low as 0 degrees C initial and higher than 300 degrees C final inlet temperatures produced the relatively high responses. Two mass spectrometric parameters were also evaluated. Indoor dust, lake sediment and human placenta tissue samples were analyzed to demonstrate reliability and sensitivity improvement of the PTV large volume injection.
More Related Videos
05:31Gas Chromatography-Mass Spectrometry Paired with Total Vaporization Solid-Phase Microextraction as a Forensic Tool
Published on: May 25, 2021
05:20Determination of 45 Pesticides in Avocado Varieties by the QuEChERS Method and Gas Chromatography-Tandem Mass Spectrometry
Published on: December 8, 2023
Related Concept Videos
Gas Chromatography: Types of Detectors-II
Gas Chromatography: Sample Injection Systems
Two primary injection methods are used...
Gas Chromatography–Mass Spectrometry (GC–MS)
A gas chromatograph consists of a long, narrow capillary column with a polysiloxane coating on the inner wall. The coating...
Gas Chromatography: Types of Columns and Stationary Phases
For an analyte to remain on the column for a sufficient amount of time, it must exhibit some level of compatibility (or...
Gas Chromatography: Types of Detectors-I
TCD is the earliest and most widely used detector that operates by measuring the changes in the thermal conductivity of the carrier gas. When a sample compound enters the detector,...
Gas Chromatography: Introduction
In GC, a sample is vaporized and mixed with an inert carrier gas (the mobile phase), which transports it through a column.