Partial volume correction strategies for quantitative FDG PET in oncology
Nikie J Hoetjes1, Floris H P van Velden, Otto S Hoekstra
1Department of Nuclear Medicine & PET Research, VU University Medical Center, Amsterdam, The Netherlands. n.hoetjes@vumc.nl
Partial volume correction (PVC) in positron emission tomography (PET) improves accuracy of standardized uptake value (SUV) without significantly impacting test-retest variability. Reconstruction-based PVC offers superior precision, while image-based methods provide practical alternatives for clinical use.
Area of Science:
- Medical imaging
- Nuclear medicine
- Quantitative PET imaging
Background:
- Quantitative accuracy in PET is compromised by partial volume effects.
- Decreasing tumor volume leads to underestimation of standardized uptake value (SUV).
- Partial volume correction (PVC) methods aim to improve SUV accuracy.
Purpose of the Study:
- To assess the accuracy and precision of different PVC methods.
- To evaluate PVC performance in simulations, phantom experiments, and clinical studies.
- To determine the impact of PVC on SUV measurements and treatment response assessment.
Main Methods:
- Evaluated three PVC methods: reconstruction-based, iterative deconvolution, and spill-in/spill-out factors.
- Used mathematical phantoms and NEMA NU2 phantoms for simulations and experiments.
- Conducted clinical studies on non-small cell lung cancer and breast cancer patients.
Main Results:
- All PVC methods accurately recovered SUV within 10% for volumes >= 1 ml.
- Reconstruction-based PVC demonstrated superior precision compared to image-based methods.
- Clinical PVC increased SUV by 5-80%, with minor effects on test-retest variability and response assessment.
Conclusions:
- PVC enhances SUV accuracy in PET imaging.
- Reconstruction-based PVC is more precise but requires specialized software.
- Image-based PVC methods are practical alternatives with comparable clinical performance.
More Related Videos
06:53Management of Respiratory Motion Artefacts in 18F-fluorodeoxyglucose Positron Emission Tomography using an Amplitude-Based Optimal Respiratory Gating Algorithm
Published on: July 23, 2020
09:58Semi-quantitative Assessment Using [18F]FDG Tracer in Patients with Severe Brain Injury
Published on: November 9, 2018
Related Concept Videos
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body being...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET