Determination of five beta-blockers in wastewaters by coupled-column liquid chromatography and fluorescence detection
P Parrilla Vázquez1, M Martínez Galera, A Serrano Guirado
1Department of Analytical Chemistry, University of Almería, La Cañada de San Urbano, 04120 Almería, Spain. parrilla@ual.es
Abstract:
A simple multidimensional system for direct injection of large volumes has been developed for the determination of five beta-blockers (atenolol, nadolol, metoprolol, bisoprolol and betaxolol) in wastewater using fluorescence detection. A C18 50 mm x 4.6 mm i.d. column coupled to a RP Amide C16 150 mm x 4.6 mm i.d. column for analyte clean-up and determination were used, respectively. The capability of a first column for eliminating large, interfering molecules, combined with an optimised, coupled-column liquid chromatography separation procedure (LC-LC), large volume injection (LVI) and fluorescence detection (FD), gave excellent sensitivity and selectivity for the target analytes. The LVI-LC-LC-FD method combines analyte isolation, preconcentration and determination into a single step. Detection limits obtained in wastewater were lower than, or equal to, 0.0020 microg L(-1). Limits of quantification (LOQs) obtained in the matrix according to IUPAC, ranged between 0.0052 and 0.0089 microg L(-1), whereas LOQs calculated according to EURACHEM Guidance, varied between 0.4 and 0.6 microg L(-1). Accuracy values ranged from 82 to 107% (n=3) and relative standard deviation (RSD) values ranged from 0.8 to 9%. The LVI-LC-LC-FD method was applied for determining the target analytes in wastewater samples obtained in Almería (Spain).
More Related Videos
05:46Identification of Pharmaceuticals in The Aquatic Environment Using HPLC-ESI-Q-TOF-MS and Elimination of Erythromycin Through Photo-Induced Degradation
Published on: August 1, 2018
06:25Post Column Derivatization Using Reaction Flow High Performance Liquid Chromatography Columns
Published on: April 26, 2016
Related Concept Videos
High-Performance Liquid Chromatography: Types of Detectors
Supercritical Fluid Chromatography
SFC utilizes a supercritical fluid mobile phase,...
High-Performance Liquid Chromatography: Instrumentation
High-Performance Liquid Chromatography: Introduction
In HPLC, two phases play a critical role in the separation process:
High-Performance Liquid Chromatography: Elution Process
Gas Chromatography: Types of Detectors-I
TCD is the earliest and most widely used detector that operates by measuring the changes in the thermal conductivity of the carrier gas. When a sample compound enters the detector,...