Multivariate calibration of spectrophotometric data using a partial least squares with data fusion

Ling Gao1, Shouxin Ren

  • 1Department of Chemistry, Inner Mongolia University, West University Road 235, Huhehot, 010021 Inner Mongolia, PR China.

Summary

A new data fusion combined with partial least squares (DF-PLS) method accurately determines multiple metal ions simultaneously. This advanced technique significantly improves spectral analysis accuracy compared to traditional partial least squares (PLS).

Related Concept Videos

Calibration Curves: Linear Least Squares01:20

Calibration Curves: Linear Least Squares

A calibration curve is a plot of the instrument's response against a series of known concentrations of a substance. This curve is used to set the instrument response levels, using the substance and its concentrations as standards. Alternatively, or additionally, an equation is fitted to the calibration curve plot and subsequently used to calculate the unknown concentrations of other samples reliably.
For data that follow a straight line, the standard method for fitting is the linear...
Instrument Calibration01:12

Instrument Calibration

Instrument calibration is essential for ensuring that instruments produce accurate and consistent results. It is vital in manufacturing, healthcare, testing laboratories, and scientific research. Calibration processes are specific to each instrument and help enhance data accuracy. Each instrument has a unique calibration process tailored to its design and function to improve data accuracy.
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...
Spectrophotometry: Introduction01:16

Spectrophotometry: Introduction

Spectrophotometry is the quantitative measurement of the absorption, reflection, diffraction, or transmission of electromagnetic radiation through a material as a function of the intensity and wavelength of the radiation. A spectrophotometer is a device used to measure the change in the radiation intensity caused by its interaction with the material.
The essential components of a spectrophotometer include a source of electromagnetic radiation, a slot for placing a material to be analyzed, and a...
Calibration Curves: Correlation Coefficient01:10

Calibration Curves: Correlation Coefficient

In a linear calibration curve, there is a value called the calibration coefficient, denoted by 'r,' which measures the strength and the direction of association between two variables. The correlation coefficient value ranges from −1 to +1. A value of +1 indicates a perfect positive linear correlation, −1 denotes a perfect negative correlation, and 0 implies no correlation between the two variables. A positive correlation value establishes that as one variable increases, the other increases, and...
UV–Vis Spectroscopy: Beer–Lambert Law01:09

UV–Vis Spectroscopy: Beer–Lambert Law

The Beer-Lambert law describes the relationship between absorbance and concentration, which combines the principles established by scientists Johann Heinrich Lambert and August Beer. Lambert's law states that when light passes through a medium, the loss in intensity is directly proportional to the original intensity and the path length of the light. Beer's law proposed that the transmittance of a solution remains constant if the product of concentration and path length is constant. The modern...
UV–Vis Spectrometers01:14

UV–Vis Spectrometers

The absorbance of UV and visible (UV–visible) radiations is measured using a UV–visible spectrophotometer. Deuterium lamps, which emit UV radiation, and tungsten lamps, which produce radiation in the visible region, are used as light sources in UV–visible spectrophotometers. A monochromator or prism is used for diffraction grating, i.e., to split the incoming radiation into different wavelengths. A system of slits is used to focus the desired wavelength on the sample cell. Samples for...