Differentiation and quantification of C1 and C2 (13)C-labeled glucose by tandem mass spectrometry
Wei Jiang1, Vicki H Wysocki, Eric D Dodds
1Department of Chemistry and Biochemistry, University of Arizona, Tucson, 85721, USA.
Abstract:
The fragmentation patterns of various (13)C-labeled glucose molecules were analyzed by electrospray ionization tandem mass spectrometry. Derivatization of glucose to yield methylglucosamine makes the C-C bond between C1 and C2 a favored cleavage site. This is in contrast to underivatized glucose, which favorably undergoes loss of a fragment containing both C1 and C2. Based on the fragmentation pattern of methylglucoasmine, we developed a method to distinguish and quantify C1 and C2 (13)C-labeled glucose by derivatization with methylamine followed by multiple reaction monitoring scans in a Q-trap mass spectrometer. Fragment ion ratios in the tandem mass spectra showed an isotope effect with (13)C or deuterium labeling, so a "correction factor" was introduced to make the quantification more accurate. The current approach can be applied to individually monitor the metabolic origin and fate of C1 and C2 atoms in (13)C-labeled glucose. This method provides a new means of quantifying glucose isotopomers in metabolic studies.
More Related Videos
08:56Quantitative Analysis of the Cellular Lipidome of Saccharomyces Cerevisiae Using Liquid Chromatography Coupled with Tandem Mass Spectrometry
Published on: March 8, 2020
14:42Liquid Chromatography Coupled to Refractive Index or Mass Spectrometric Detection for Metabolite Profiling in Lysate-based Cell-free Systems
Published on: September 23, 2021
Related Concept Videos
Tandem Mass Spectrometry
Peptide Identification Using Tandem Mass Spectrometry
This technique helps gather information regarding the protein from which the peptide was obtained and to study the peptides’ amino acid sequence. Identifying peptides from a complex mixture is an important component of the growing field of...
Mass Spectrometry: Complex Analysis
GC–MS is a powerful hyphenated method commonly used in forensics and environmental...
¹³C NMR: ¹H–¹³C Decoupling
A broadband decoupling technique is used to simplify these complex, sometimes overlapping, signals. Broadband decoupling relies on a...
MALDI-TOF Mass Spectrometry
High-Resolution Mass Spectrometry (HRMS)