Identification and control of metal-chelating chromatographic artifacts in the analysis of a malonohydrazide derivative drug compound
Related Experiment Video
Updated: Jun 13, 2026

Quantification of Metal Leaching in Immobilized Metal Affinity Chromatography
Published on: January 17, 2020
Shaoping Yang1, Anna Caltabiano, Lianming Wu
1Product Development Group, Pharmaceutical Development, GlaxoSmithKline, USA.
Abstract:
Two unusual chromatographic artifact peaks were detected in the HPLC analysis for content of a malonohydrazide derivative drug and drug-related impurities. The artifacts were identified as the copper(II) chelating complexes with the drug compound and one of the process impurities. Our investigations suggested that built-up of Cu(2+) contamination in the HPLC system was the primary source for formation of the chelating artifacts. A rinse procedure using diluted EDTA solution was developed, and demonstrated to effectively purge trace level of heavy metals including Cu(2+) from the system, and therefore inhibited the formation of both chelates. Furthermore, the rinse was shown to introduce no detrimental impact on the response accuracy of the active drug compound and related impurities.
More Related Videos
07:54Analysis of AtHIRD11 Intrinsic Disorder and Binding Towards Metal Ions by Capillary Gel Electrophoresis and Affinity Capillary Electrophoresis
Published on: August 22, 2018
09:42Fabrication of a Dipole-assisted Solid Phase Extraction Microchip for Trace Metal Analysis in Water Samples
Published on: August 7, 2016
Related Concept Videos
Extraction: Advanced Methods
Masking and Demasking Agents
There are many masking agents, such as cyanide, fluoride, triethanolamine, thiourea, and 2,3-bis(sulfanyl)propan-1-ol (formerly 2,3-dimercapto-1-propanol), with the masking agent chosen based on the metal...
Affinity Chromatography