Related Experiment Video
Updated: Jun 13, 2026

Merging Ion Concentration Polarization between Juxtaposed Ion Exchange Membranes to Block the Propagation of the Polarization Zone
Published on: February 23, 2017
Capillary-valve-based fabrication of ion-selective membrane junction for electrokinetic sample preconcentration in
Vincent Liu1, Yong-Ak Song, Jongyoon Han
1Department of Electrical Engineering and Computer Science, Massachusetts Institute of Technology, Cambridge, MA 02139, USA.
Abstract:
In this paper, we report a novel method for fabricating ion-selective membranes in poly(dimethylsiloxane) (PDMS)/glass-based microfluidic preconcentrators. Based on the concept of capillary valves, this fabrication method involves filling a lithographically patterned junction between two microchannels with an ion-selective material such as Nafion resin; subsequent curing results in a high aspect-ratio membrane for use in electrokinetic sample preconcentration. To demonstrate the concentration performance of this high-aspect-ratio, ion-selective membrane, we integrated the preconcentrator with a surface-based immunoassay for R-Phycoerythrin (RPE). Using a 1x PBS buffer system, the preconcentrator-enhanced immunoassay showed an approximately 100x improvement in sensitivity within 30 min. This is the first time that an electrokinetic microfluidic preconcentrator based on ion concentration polarization (ICP) has been used in high ionic strength buffer solutions to enhance the sensitivity of a surface-based immunoassay.
Related Concept Videos
Capillary Electrophoresis: Instrumentation
Potentiometry: Membrane Electrodes