Background:
Nasal intermittent positive pressure ventilation (NIPPV) is a mode of non-invasive respiratory support. Its mechanisms of action and optimal delivery techniques are unknown.
Aim:
This observational study aimed to investigate and quantify delivered peak pressures during non-synchronised ventilator-generated NIPPV.
Methods:
Infants born below 30 weeks gestation receiving ventilator-generated NIPPV delivered via Hudson prongs were recruited. Intraprong pressure, change in tidal volume, respiratory rate, oxygen saturations, inspired oxygen and video images were recorded.
Results:
Eleven infants (four infants were female) of median (interquartile range; IQR) gestational age 25(+/-3) (25(+/-2) 26(+/-0)) weeks and birth weight 732 (699-895) g, were studied at 24 (19-41) days of age. Six infants, with set peak pressure (peak inflation pressure; PIP) of 20 cm H(2)O, received a median pressure of 15.9 (IQR 13.6-17.9) cm H(2)O. 37% of inflations were delivered at least 5 cm H(2)O below set PIP. 12.7% of inflations were delivered above set PIP. Five infants with set PIP of 25 cm H(2)O received a median PIP of 17.2 (IQR 15.0-18.3) cm H(2)O. 83% of inflations were delivered at least 5 cm H(2)O below set PIP, with 6.1% delivered higher than set PIP. The difference in delivered PIP between the groups was 1.3 cm H(2)O. PIP was highest and most variable when the infant was moving. Delivered PIP did not vary whether it coincided with spontaneous inspiration or expiration.
Conclusion:
During ventilator-generated non-synchronised NIPPV delivered PIP was variable and frequently lower than set PIP. Delivered PIP was occasionally greater than set PIP.
Related Concept Videos
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation (NIPPV)
Mechanical Ventilation I: Indication and Settings
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...
Pulmonary Ventilation: Inhalation
Boyle's law becomes particularly pertinent when examining respiratory...
Pressure Relationships in Thoracic Cavity
Breathing Mechanisms
Both intra-alveolar and intrapleural pressures rely on specific lung properties. The ability to breathe—allowing air to enter the lungs during...