Related Experiment Video
Updated: Jun 12, 2026

Hemodynamic Precision in the Neonatal Intensive Care Unit using Targeted Neonatal Echocardiography
Published on: January 27, 2023
Initial experience with a miniaturized multiplane transesophageal probe in small infants undergoing cardiac
Sinai C Zyblewski1, Girish S Shirali, Geoffrey A Forbus
1Division of Pediatric Cardiology, Medical University of South Carolina, Charleston, South Carolina 29425, USA.
Purpose:
There has been reluctance to use intraoperative transesophageal echocardiography (TEE) in small infants. We assessed the utility and safety of a new miniaturized multiplane micro-TEE probe in small infants undergoing cardiac operations.
Description:
Hemodynamic and ventilation variables were prospectively recorded before and after micro-TEE insertion and removal in infants weighing 5 kg or less undergoing cardiac operations.
Evaluation:
The study included 42 patients with a mean weight of 3.6 +/- 0.9 kg (range, 1.7 to 5 kg). All probe insertions were successful. There were no complications or clinically significant changes in hemodynamic or ventilation variables. Information provided by TEE resulted in surgical revision in 6 of the 42 patients.
Conclusions:
The micro-TEE provides high quality, useful diagnostic images without hemodynamic or ventilation compromise in small infants undergoing cardiac operations. This advance is important with the growing trend towards complete repair of complex structural heart disease in small infants.