Related Experiment Video
Updated: Aug 22, 2026

Construction of a Preclinical Multimodality Phantom Using Tissue-mimicking Materials for Quality Assurance in Tumor Size Measurement
Published on: July 29, 2013
Comparison of Four CBCT-IGRT Geometric QA Phantoms within a Single Institutional Workflow
Hyojun Park1,2, Hyejo Ryu1, Jin Hwa Choi3
1Department of Radiation Oncology, Chung-Ang University Gwangmyeong Hospital, Gyeonggi-do, 14353, Republic of Korea.
Objectives:
To compare four commercial cone-beam CT (CBCT) geometric quality assurance phantoms within a common workflow, focusing on positioning/repositioning accuracy and agreement between the laser-defined setup reference and the CBCT imaging isocentre.
Methods:
MIMI™, QUASAR™ Penta-Guide, ISOCube, and MZ CUBE were evaluated monthly for five months on a Versa HD system with XVI CBCT and a HexaPOD couch. Two observers independently completed each measurement. Three-dimensional post-correction translational residuals (Δdpos/rep) and laser-defined setup reference-CBCT imaging isocentre distances (Δdiso) were calculated. Phantom comparisons used observer-averaged values from the five monthly sessions. Rotational correction and CBCT image quality were assessed descriptively with MZ CUBE.
Results:
Mean Δdpos/rep values ranged from 0.19 to 0.31 mm, with mean absolute observer differences of 0.05-0.12 mm. No overall difference was found among the four phantoms (p = 0.547). Mean Δdiso values ranged from 0.68 to 0.87 mm, and all measurements were below 2 mm. An overall difference was observed for Δdiso (p = 0.019), although no pairwise comparisons were performed. With MZ CUBE, mean post-correction translational and rotational residuals were 0.24 mm and 0.37°, respectively, and image-quality measures showed no consistent temporal trend.
Conclusions:
The four phantoms showed similarly small post-correction translational residuals. In our setting, MZ CUBE additionally supported rotational and image-quality checks.
Advances In Knowledge:
Direct comparison within a common workflow showed similar translational performance, while the scope of additional testing depended on the phantom configuration and locally available accessories.

